Z-1468-2012

VITROS® 3600 Immunodiagnostic System

Ortho-Clinical Diagnostics, Inc. / initiated 2012-03-15 / Terminated / root cause: manufacturing process

Ortho-Clinical Diagnostics began a recall of VITROS® 3600 Immunodiagnostic System Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of... on . The reason FDA records is "Ortho-Clinical Diagnostics is recalling VITROS Chemistry Systems due to possible internal hard drive failure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1468-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS® 3600 Immunodiagnostic System Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Ortho-Clinical Diagnostics is recalling VITROS Chemistry Systems due to possible internal hard drive failure.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Ortho Clinical Diagnostics (OCD) sent an Important Product Correction Notification letter dated March 15, 2012 via FedEx overnight courier to all affected customers. Foreign affiliates were informed by email on 15 March 2012 of the issue and instructed to notify their consignees of the issue and actions. The letter identified the affected products, problem and actions to be taken. The letter instructed customers to: 1) Consider performing the backup procedure more frequently to help prevent the loss of data on your system; 2) Contact OCD Customer Technical Service representatives if a hard drive failure occurs on your system; A Field Engineer will be dispatched to replace the hard drive on your system; 3) Complete and return the Confirmation of Receipt form upon receipt of this notification; and 4) Post this notification by each VITROS System in your facility or with the user documentation. For questions, refer to the Questions and Answers section. For additional questions, contact Customer Technical Services at 1-800-421-3311.
Quantity in commerce
173
Distribution
Worldwide Distribution
Product codes and lots
Serial number range: 36000467 - 36000639