VITROS® 5600 Chemistry System
Ortho-Clinical Diagnostics began a recall of VITROS® 5600 Chemistry System, Product Code: 6802413 Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a... on . The reason FDA records is "Ortho-Clinical Diagnostics is recalling VITROS Chemistry Systems due to possible internal hard drive failure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1470-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS® 5600 Chemistry System, Product Code: 6802413 Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Ortho-Clinical Diagnostics is recalling VITROS Chemistry Systems due to possible internal hard drive failure.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Ortho Clinical Diagnostics (OCD) sent an Important Product Correction Notification letter dated March 15, 2012 via FedEx overnight courier to all affected customers. Foreign affiliates were informed by email on 15 March 2012 of the issue and instructed to notify their consignees of the issue and actions. The letter identified the affected products, problem and actions to be taken. The letter instructed customers to: 1) Consider performing the backup procedure more frequently to help prevent the loss of data on your system; 2) Contact OCD Customer Technical Service representatives if a hard drive failure occurs on your system; A Field Engineer will be dispatched to replace the hard drive on your system; 3) Complete and return the Confirmation of Receipt form upon receipt of this notification; and 4) Post this notification by each VITROS System in your facility or with the user documentation. For questions, refer to the Questions and Answers section. For additional questions, contact Customer Technical Services at 1-800-421-3311.
- Quantity in commerce
- 417
- Distribution
- Worldwide Distribution
- Product codes and lots
- Serial number range: 56001033 - 56001449