Coloplast, Restorelle DirectFix, Posterior Polypropylene Mesh, Catalogue Number 501460, Rx Only...
Coloplast Manufacturing US, LLC began a recall of Coloplast, Restorelle DirectFix, Posterior Polypropylene Mesh, Catalogue Number 501460, Rx Only, Manufacturer, Coloplast A/S 3050 Denmark Restorelle is a... on . The reason FDA records is "Coloplast is conducting a field correction based on the conclusion that two lots of Restorelle DirectFix Posterior mesh were mislabeled. The incorrect size was printed on the patient label. Specifically the patient...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1481-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Coloplast, Restorelle DirectFix, Posterior Polypropylene Mesh, Catalogue Number 501460, Rx Only, Manufacturer, Coloplast A/S 3050 Denmark Restorelle is a ploypropylene mesh that may be used for the repair of abdominal wall hernia, including inguinal, fermoral, and incisional, and uterovaginal proplapse and other fascial deficiencies that require support material. It may be used in open or laparoscopic abdominal procedures or for repair by the vaginal route.
- Product code
- FTL Mesh, Surgical, Polymeric
- Reason for recall
- Coloplast is conducting a field correction based on the conclusion that two lots of Restorelle DirectFix Posterior mesh were mislabeled. The incorrect size was printed on the patient label. Specifically the patient label depicts the size as 16x11 cm, however, the correct size is actually 14x15 cm. It is important to note that both the product and box labels were correct.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Coloplast sent a notification letter dated March 2, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were informed their coloplast representative is providing replacement device (s) and new patient labels. For further questions please call 1 (800) 258-3476.
- Quantity in commerce
- 134
- Distribution
- USA ( nationwide) including the states of AL, AZ, AR, CA, CO, FL, GA, ID, IL, KY, MD, MS, MO, NY, PA, TN, VA,WA and WI.
- Product codes and lots
- Lot numbers : 2936642, 2967400.