Z-1397-2012

Sterile surgical kits containing the Umbilicup device: DeRoyal (R) VAGINAL DELIVERY PACK, REF 89-5260.04, Rx...

DeRoyal Industries, Inc. / initiated 2012-01-20 / Terminated / root cause: manufacturing process

DeRoyal Industries Inc began a recall of Sterile surgical kits containing the Umbilicup device: DeRoyal (R) VAGINAL DELIVERY PACK, REF 89-5260.04, Rx Only, STERILE EO, Manufacturer: DeRoyal, Powell... on . The reason FDA records is "Device's needle may become dislodged from the cup during shipment or during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1397-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
DeRoyal Industries, Inc.
Product
Sterile surgical kits containing the Umbilicup device: DeRoyal (R) VAGINAL DELIVERY PACK, REF 89-5260.04, Rx Only, STERILE EO, Manufacturer: DeRoyal, Powell, TN 37849 DeRoyal (R) VAGINAL DELIVERY TRAY, REF 89-3928.10, Rx Only, STERILE EO, Manufacturer: DeRoyal, Powell, TN 37849 DeRoyal (R) C-SECTION TRAY, REF 89-2960.20, Rx Only, STERILE EO, Manufacturer: DeRoyal, Powell, TN 37849 Product Usage: Collection of cord blood
Product code
LRP Tray, Surgical
Reason for recall
Device's needle may become dislodged from the cup during shipment or during use.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
DeRoyal Industries sent Recall communication by letter to all affected consignees of the Non-Sterile Bulk and Single Sterile Units on January 20, 2012 and of the Kits and Trays on Febraury 6, 2012. The letters identified the affected products, problem and actions to be followed. Customers were instrructed to complete the enclosed Notice of Destruction form and return this form by fax (865-362-3716) or email ([email protected]) within two weeks of the date of this letter. DeRoyal will use your response on the Notice of Destruction form to send replacements as soon as they become available. Distributors were instructed to forward this recall to any of their end users that may have affected product. If you have questions or need assistance with the recall, please call 865.362.1020 or Robby Lockett at 865.362.1034.
Quantity in commerce
1562 units
Distribution
Worldwide Distribution - USA Nationwide and the countries of: Canada, Dominican Republic, South Africa, and the United Arab Emirates.
Product codes and lots
REF 89-5260.04, Lot Numbers: 17636582, 18080838, 18146577, 18178608, 18298693, 18952510, 19419055, 19480569, 19518336, 19592131, 19901141, 20128925, 20176492, 20678329, 20724372, 20822626, 21815858, 21923306, 22732536, 24157900, 24291754, 24630752, 24747336, 24823319, 25460353, 25721453, 25784489, 26018591, 26037952, 26149090, 26280549, 27163133, 27206527; REF 89-3928.10, Lot Numbers: 17826968, 17857924, 18177242, 18407790, 19021830, 19140417, 19676836, 20009408, 20127383, 20495653, 20676711, 20887981, 21023367, 21184339, 21531020, 21717131, 21839219, 22016935, 22057411, 22204397, 22379218, 22585169, 22731293, 23094151, 23420711, 23874561, 24230804, 24429928, 24519157, 24774801, 26098270, 26316468, 26663347, 26728235, 26956844, 27180961, 27343071, 27826304, 27911550, 27931729; REF 89-2960.20, Lot Numbers: 17903599, 17935806, 18183757, 19140610, 19325542, 19480673, 19881274, 20208693, 20679330, 20822829, 21056863, 21067909, 21318966, 21422472, 21649319, 21816375, 22123933, 22587164, 22782189, 23422927, 23696522, 23949622, 24007645, 24073423, 24232105, 24432131, 24468732, 24576258, 24823474, 26019527, 26281120, 26368444, 26785141, 26923452, 27227192, 27473916, 27931868