Straumann Standard Plus Implant 04.8mm Wide Neck SLA 10mm Article Number: 043.652S. Dental implants are...
Straumann USA, LLC began a recall of Straumann Standard Plus Implant 04.8mm Wide Neck SLA 10mm Article Number: 043.652S. Dental implants are intended for surgical placement in maxillary and/or... on . The reason FDA records is "Mislabeled: Vial lid label on some of the dental implant vials from this lot incorrectly state 04.8 SP 10 RN instead of 04.8 SP 10 WN". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1351-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Straumann USA, LLC
- Product
- Straumann Standard Plus Implant 04.8mm Wide Neck SLA 10mm Article Number: 043.652S. Dental implants are intended for surgical placement in maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous or partially edentulous patients. Dental implants are for single-stage or two-stage surgery. Dental implants are intended for immediate placement and function on single-tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function. Multiple tooth applications may be rigidly splinted. In the case of edentulous patients, four or more implants must be used.
- Product code
- NDP Accessories, Implant, Dental, Endosseous
- Reason for recall
- Mislabeled: Vial lid label on some of the dental implant vials from this lot incorrectly state 04.8 SP 10 RN instead of 04.8 SP 10 WN
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Straumann sent a "FIELD CORRECTIVE ACTION NOTICE" dated March 12, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The letter instructs customers to check their inventory for the affected product and to return any affected product to the firm. Additionally, a Customer Confirmation Form was enclosed for customers to complete and return. Contact the firm at 978-747-2514 for questions regarding this notice.
- Quantity in commerce
- 346 units
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Lot No. CW642