Z-1643-2012

BD Focal Point"230V LC (Refurbished), Catalog number: 490681. The BD Focal Point" Slide Profiler (formerly...

Becton Dickinson & Co / initiated 2011-10-06 / Terminated

Tripath Imaging, Inc. began a recall of BD Focal Point"230V LC (Refurbished), Catalog number: 490681. The BD Focal Point" Slide Profiler (formerly the AutoPap System) is an automated cervical... on . The reason FDA records is "The cable tie mounts that adhere to the inside of the FocalPoint do not adhere as intended and can therefore become detached". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1643-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Focal Point"230V LC (Refurbished), Catalog number: 490681. The BD Focal Point" Slide Profiler (formerly the AutoPap System) is an automated cervical cytology screening device intended for use in initial screening of cervical cytology slide.
Product code
MKQ Processor, Cervical Cytology Slide, Automated
Reason for recall
The cable tie mounts that adhere to the inside of the FocalPoint do not adhere as intended and can therefore become detached.
FDA root cause
Component change control
Action
BD Diagnostics first notified consignees by letter on October 6, 2011. BD Diagnostics also sent a letter dated February 2012, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Field Representatives were advised to inspect all cables, adhesive cable ties and connections to ensure they are secure. Inspections and repairs, if necessary, would be carried out at the next service call for each instrument, or no later than the next regularly scheduled preventative maintenance visit. For questions consignees were instructed to contact BD Diagnostics Technical Support at 1-877-822-7771 option 2 in the US, their local distributor, or their local BD office for any questions. For questions regarding this recall call 918-563-1624.
Quantity in commerce
see line 1.
Distribution
Worldwide Distribution - USA including AZ, CT, IL, IN, FL, MN, NC, OK, OR, and TX and the countries of Australia, Belgium, Brazil, Canada, Japan and Thailand
Product codes and lots
To be obtained