Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Systems. Brilliance 64 Model# 728231, Ingenuity CT...
Philips Medical Systems (Cleveland) Inc began a recall of Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Systems. Brilliance 64 Model# 728231, Ingenuity CT Model # 728326, Distributed By: Philips Healthcare... on . The reason FDA records is "Philips Healthcare is performing an upgrade to the field for issues related to Philips Brilliance 64 and Ingenuity CT scanner having Software Version 3.5.1. When the user plans a scan with tilt applied, the scan...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1361-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Systems. Brilliance 64 Model# 728231, Ingenuity CT Model # 728326, Distributed By: Philips Healthcare, Cleveland, OH. The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Philips Healthcare is performing an upgrade to the field for issues related to Philips Brilliance 64 and Ingenuity CT scanner having Software Version 3.5.1. When the user plans a scan with tilt applied, the scan increment changes from the user defined values (noncontiguous) to a contiguous scan.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- .
- Quantity in commerce
- 440 units
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AK, AR, AZ, CA, CO, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MT, ND, NE, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA and WV, and the countries of Australia, Bahamas, Bangladesh, Belgium, Bolivia, Brazil, Chile, China, Colombia, Croatia, Denmark, Finland, France, Georgia, Germany, Great Britain, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Laos, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Poland, Russia, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom and Vietnam.
- Product codes and lots
- System Code #728231, 4003, 4021, 9033, 9037, 9041, 9045, 9048, 9062, 9064, 9070, 9075, 9080, 9108, 9120, 9142, 9151, 9196, 9197, 9201, 9202, 9205, 9206, 9208, 9214-9218, 9462, 9508, 9531, 9534, 9535, 9544, 9554, 9561, 9584, 9603, 9606, 9608, 9613, 9627, 9630, 9657, 9668, 9671, 9672, 9695, 9706, 9708, 9719, 9728, 9734, 9735, 9741, 9745, 9756, 9789, 9819, 9832, 9838, 9839, 9840, 9859, 9874, 9875, 9891, 9898, 9904, 9918, 9926, 9952, 9964, 9965, 9968, 9975, 9997, 9999, 10000, 10008, 10015, 10041, 10052, 10054, 10092, 10095, 10100, 10103, 10114, 10125, 10145, 10151, 10152, 10154, 10170-10172, 10192, 10196, 10198, 10221, 10222, 10251, 10256, 10273, 10286, 10294, 10307, 10312, 10336, 10337, 10350, 10352, 10362, 10363, 10372, 10376, 10380, 10386, 10393, 10412, 10425, 10426, 10428, 10438, 10445, 10446, 10452, 10459, 10461, 10468, 10470, 10471, 10473, 10482, 10486, 10489, 10491, 10493, 10497, 10503, 10505, 10511-105531, 10533-10555, 10557-10566, 10568-10575, 10577-10604, 10606, 10608-10620, 10622-10624, 10628-10640, 29037, 29049, 29057, 64040, 90036, 90054, 90058, 90063, 90080, 90100, 90102, 90106, 90117, 90135, 90169, 90182, 91003, 95015, 95055, 95065, 95074, 95077, 95101, 95108, 95130, 95133, 95138, 95140, 95157, 95168, 95174, 95178, 95184, 95185, 95191, 95207, 95243, 95295, 95315, 95324, 95327, 95350, 95357, 95358, 95362, 95389, 95403, 95414, 95425, 95429, 95443, 95455, 95504, 95516, 95529, 95534, 95568, 95570, 95572, 95573, 95575, 95602, 95605, 95609, 95613, 95624, 95630, 95638, 95652, 95659, 95664, 95665, 95677, 95684, 95687, 95697, 95700-95728 & 300074. System Code #: 728326, Serial #: 9511, 300003-300005, 300010-300022, 300024, 300025, 300027-300066 & 300068-300073.