Dimension Vista(R) MG Flex(R) Reagent Cartridge The Dimension Vista(R) MG method is an in vitro diagnostic...
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista(R) MG Flex(R) Reagent Cartridge The Dimension Vista(R) MG method is an in vitro diagnostic test for the quantitative measurement of magnesium... on . The reason FDA records is "Firm has confirmed a negative shift of approximately 0.2mg/dL [0.08 mmol/L] in patient and QC samples when using Vista Magnesium". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1433-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista(R) MG Flex(R) Reagent Cartridge The Dimension Vista(R) MG method is an in vitro diagnostic test for the quantitative measurement of magnesium in human serum, plasma and urine on the Dimension Vista(R) System
- Product code
- JGJ Photometric Method, Magnesium
- Reason for recall
- Firm has confirmed a negative shift of approximately 0.2mg/dL [0.08 mmol/L] in patient and QC samples when using Vista Magnesium
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens sent an Urgent Field Safety Notification letter dated February 5, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The letter explained the correction needed and instructed customers to forward the letter to anyone they may have distributed the affected product. For questions or concerns call 800-441--9250. For questions regarding this recall call 301-631-6299.
- Quantity in commerce
- 9,178
- Distribution
- Nationwide Distribution including Puerto Rico and within the US to Washington, DC and the following States: AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Puerto Rico.
- Product codes and lots
- Lot numbers 11263AC, 11286AB, 11306BA, 11332AB, 11332AC and 12010AB.