Z-1383-2012

Siemens Axiom Luminos dRF solid state x-ray imager Solid state x-ray imager (flat panel/dig. imager)

Siemens Medical Solutions USA, Inc. / initiated 2012-02-16 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Siemens Axiom Luminos dRF solid state x-ray imager Solid state x-ray imager (flat panel/dig. imager) on . The reason FDA records is "During regular product monitoring, Siemens became aware of a potential issue with the AXIOM Luminos dRF. Firm became aware of a potential issue with torque wrenches - during a regular scheduled calibration of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1383-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Axiom Luminos dRF solid state x-ray imager Solid state x-ray imager (flat panel/dig. imager)
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
During regular product monitoring, Siemens became aware of a potential issue with the AXIOM Luminos dRF. Firm became aware of a potential issue with torque wrenches - during a regular scheduled calibration of the wrench, it was discovered that the wrench was out of the specification. The wrench was over-torquing approximately 21% for low-end readings (50 ftlbs) and 9% for high-end readings (250
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens initiated a recall/correction on February 16, 2012 by issuing Update Instruction XP056/11/S to the affected customers. All screws on the unit support on AXIOM Luminos dRF units will be replaced by a Siemens service engineer. For information regarding this recall call 610-219-4834.
Quantity in commerce
2
Distribution
Nationwide Distribution including California
Product codes and lots
Model number 10094200, serial numbers 3019 and 1358