Non-DEHP Administration Set with Non-Vented Spike and 1.2 Micron Filter and Check Valve Packaged Assembly The...
MOOG Medical Devices Group began a recall of Non-DEHP Administration Set with Non-Vented Spike and 1.2 Micron Filter and Check Valve Packaged Assembly The Curlin Ambulatory Infusion Pump System with... on . The reason FDA records is "Ambulatory administration sets used with the Curlin Ambulatory Pumps including the 4000 CMS Ambulatory Pump, 6000 CMS Ambulatory Pump and the Pain Smart IOD Ambulatory Pump may leak during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1406-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zevex, Inc.
- Product
- Non-DEHP Administration Set with Non-Vented Spike and 1.2 Micron Filter and Check Valve Packaged Assembly The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.
- Product code
- FRN Pump, Infusion
- Reason for recall
- Ambulatory administration sets used with the Curlin Ambulatory Pumps including the 4000 CMS Ambulatory Pump, 6000 CMS Ambulatory Pump and the Pain Smart IOD Ambulatory Pump may leak during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Moog Medical Device Group sent a Medical Device Recall Notification letter dated March 20, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to remove the affected product from their inventory, contact the distributor or Moog Customer Service at 1-800-970-2337, prompt #7, then return product to distributor or Moog. For additional questions regarding this recall call Moog's Customer Service Department at (800) 970-2337.
- Quantity in commerce
- 4341
- Distribution
- Worldwide Distribution - USA including AZ,,, CA,,, FL,,, GA,,, KS,,, IL,,, MA,,, MD,,, MI,,, OH,,, PA,,, TX,,, VA,,, WI,,, and the countries of Germany,,, Italy,,, Morocco,,, New Zealand,,, and the United Arab Emirates,,,
- Product codes and lots
- Product Number: 340-4128v Lot Numbers: CF1120212,,, CF1120213,,, CF1120214,,, CF1122239,,, CF1123012,,, CF1123014,,, CF1125502,,, CF1125503,,, CF1127113,,, CF1127912,,, CF1129907,,, CF1129908