RF IgA Calibrator, a component of the EL-RF/3 Rheumatoid Factor Test Kit; an in vitro diagnostic; TheraTest...
Thera Test Laboratories, Inc. began a recall of RF IgA Calibrator, a component of the EL-RF/3 Rheumatoid Factor Test Kit; an in vitro diagnostic; TheraTest Laboratories, Lombard, IL 60148; component catalog... on . The reason FDA records is "The RF IgA Calibrator component of the EL-RF/3 Rheumatoid Factor test system may be unstable, resulting in a drop in the RF IgA calibrator OD value which could invalidate an RF IgA test run". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1546-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Theratest Laboratories, Inc.
- Product
- RF IgA Calibrator, a component of the EL-RF/3 Rheumatoid Factor Test Kit; an in vitro diagnostic; TheraTest Laboratories, Lombard, IL 60148; component catalog #12833, kit catalog #303-305. The TheraTest EL-RF/3 is intended as an aid to the diagnosis of Rheumatoid Arthritis (RA).
- Product code
- DHR System, Test, Rheumatoid Factor
- Reason for recall
- The RF IgA Calibrator component of the EL-RF/3 Rheumatoid Factor test system may be unstable, resulting in a drop in the RF IgA calibrator OD value which could invalidate an RF IgA test run.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- TheraTest sent "Urgent Device Recall" letters dated March 28, 2012 to customers that received the affected kit, requesting them to return or destroy the remainder of the product. TheraTest issued a replacement kit to all customers that had unused product as well as any material they consumed which resulted in a failed run. Customers were also instructed to complete the enclosed confirmation form and fax the completed form back to TheraTest. Any questions were directed to TheraTest at 800-441-0771.
- Quantity in commerce
- 207 kits
- Distribution
- Nationwide Distribution (US)
- Product codes and lots
- RF IgA Calibrator: component catalog number 12833, lot number 11118586, exp. 28 DEC 2012; EL-RF/3: kit catalog number 303-305, lot number 11118583, exp. 28 NOV 2012.