Alere Triage Profiler SOB Panel, PN 97300. The test is used as an aid in the diagnosis of myocardial...
Alere San Diego began a recall of Alere Triage Profiler SOB Panel, PN 97300. The test is used as an aid in the diagnosis of myocardial infarction (injury), an aid in the diagnosis and... on . The reason FDA records is "Alere San Diego, Inc. is initiating a voluntary recall of the Alere Triage CardioProfiler Panel PN 9700CP, Lot W49569 and the Alere Triage Profiler SOB Panel PN 97300, Lot W48990. Alere San Diego, Inc. has determined...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1467-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alere San Diego, Inc.
- Product
- Alere Triage Profiler SOB Panel, PN 97300. The test is used as an aid in the diagnosis of myocardial infarction (injury), an aid in the diagnosis and assessment of severity of heart failure, an aid in the risk stratification of patients with heart failure, an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic events including pulmonary embolism and an aid in the risk stratification of patients with acute coronary syndromes.
- Product code
- MMI Immunoassay Method, Troponin Subunit
- Reason for recall
- Alere San Diego, Inc. is initiating a voluntary recall of the Alere Triage CardioProfiler Panel PN 9700CP, Lot W49569 and the Alere Triage Profiler SOB Panel PN 97300, Lot W48990. Alere San Diego, Inc. has determined that these lots have an increased frequency of Troponin I results >0.05ng/mL for samples which are found to be below 0.05ng/mL upon additional testing.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- A recall communication was initiated on 3/15/12 with Alere San Diego, Inc. forwarding a "Urgent Medical Device Recall" letter to all their customers who purchased the Alere Triage CardioProfiler Panel PN 97100CP, Lot W49569 and Alere Triage Profiler SOB Panel PN 97300, Lot W48990. The letter informed the customers about the problems identified and the actions to be taken. Customers with any questions about the information contained in this notification should contact Alere San Diego, Inc. at (877) 308-8287.
- Quantity in commerce
- 1,316 kits
- Distribution
- Worldwide Distribution, including Nationwide (US) and the countries of Brazil, China, India and Kuwait.
- Product codes and lots
- Lot Code: W48990 Phase 2 (June 12, 2012): Lot#W50019, W50027