Siemens Axiom Sensis XP diagnostic programmable computer Diagnostic programmable computer
Siemens Medical Solutions USA, Inc began a recall of Siemens Axiom Sensis XP diagnostic programmable computer Diagnostic programmable computer on . The reason FDA records is "There is a potential malfunction in the connection of a Sensis EP system with Carto(R) 3 dedicated cable set (part number 14411343) with the Carto 3 system from Bionsense Webster". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1506-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Axiom Sensis XP diagnostic programmable computer Diagnostic programmable computer
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- There is a potential malfunction in the connection of a Sensis EP system with Carto(R) 3 dedicated cable set (part number 14411343) with the Carto 3 system from Bionsense Webster.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens sent a recall/correction letter dated February 22, 2012, via Update Instruction AX049/11/S of this product to all affected customers. The letter informed customers of the potential issues and provides additional information on the implemented resolution. An Update Instuction AX048/11/S will be available immediately. The update instruction provides a new version of the Carto 3 addendum to the Operator Manual and the exchange of the previous stim cables with the latest cable. For questions call 610-219-4834.
- Quantity in commerce
- 3
- Distribution
- Nationwide Distribution including FL, NC, and SC.
- Product codes and lots
- Model number 10608285 - serial numbers 41021, 10006, and 6045.