Z-1464-2012
Philips Digital Diagnost Systems with Eleva software version 2.x and stitching option
Philips North America LLC / initiated 2012-04-03 / Terminated / root cause: labeling and UDI
Philips Healthcare Inc. began a recall of Philips Digital Diagnost Systems with Eleva software version 2.x and stitching option Intended use: Stationary radiographic system, Radiographic system, digital on . The reason FDA records is "Potential misdiagnosis due to improperly automatically stitched images, improper image ruler handling". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1464-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Digital Diagnost Systems with Eleva software version 2.x and stitching option Intended use: Stationary radiographic system, Radiographic system, digital
- Product code
- MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Reason for recall
- Potential misdiagnosis due to improperly automatically stitched images, improper image ruler handling.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- The firm, Philips Healthcare Imaging, issued an "URGENT-FIELD SAFETY NOTICE" dated March 30, 2012 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to follow the instructions carefully and to not use the Philips Image Ruler for measurements. The firm plans to supply an Addendum to the Instructions For Use and a Philips Service Engineer will contact the customer as soon as the Field Action Kit is ready to be implemented. If you need any further information or support concerning this issue, please contact your local Philips representative or call 1-866-767-2822.
- Quantity in commerce
- 606 units
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech. Rep, Denmark, Finland, France, Germany, Guadeloupe, India, Indonesia, Iran, Iraq, Italy, Japan, Malta, Mexico, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Salkowski, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Ver.Arab.Emir.
- Product codes and lots
- Model #s 712020, 712022, 712082 All systems with Eleva software version 2.x and stitching option.