Z-1464-2012

Philips Digital Diagnost Systems with Eleva software version 2.x and stitching option

Philips North America LLC / initiated 2012-04-03 / Terminated / root cause: labeling and UDI

Philips Healthcare Inc. began a recall of Philips Digital Diagnost Systems with Eleva software version 2.x and stitching option Intended use: Stationary radiographic system, Radiographic system, digital on . The reason FDA records is "Potential misdiagnosis due to improperly automatically stitched images, improper image ruler handling". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1464-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Digital Diagnost Systems with Eleva software version 2.x and stitching option Intended use: Stationary radiographic system, Radiographic system, digital
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
Potential misdiagnosis due to improperly automatically stitched images, improper image ruler handling.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
The firm, Philips Healthcare Imaging, issued an "URGENT-FIELD SAFETY NOTICE" dated March 30, 2012 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to follow the instructions carefully and to not use the Philips Image Ruler for measurements. The firm plans to supply an Addendum to the Instructions For Use and a Philips Service Engineer will contact the customer as soon as the Field Action Kit is ready to be implemented. If you need any further information or support concerning this issue, please contact your local Philips representative or call 1-866-767-2822.
Quantity in commerce
606 units
Distribution
Worldwide distribution: USA (nationwide) and countries of: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech. Rep, Denmark, Finland, France, Germany, Guadeloupe, India, Indonesia, Iran, Iraq, Italy, Japan, Malta, Mexico, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Salkowski, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Ver.Arab.Emir.
Product codes and lots
Model #s 712020, 712022, 712082 All systems with Eleva software version 2.x and stitching option.