Water Pik Sinusense battery operated device including a blue semi-transparent water reservoir, a white hand...
Water Pik, Inc. began a recall of Water Pik Sinusense battery operated device including a blue semi-transparent water reservoir, a white hand held button initiated injector device, and a soft... on . The reason FDA records is "Customer complaints related to cross contamination of sinus rinse solution between the battery compartment and solution reservoir used for rinsing nasal passages". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1674-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Water Pik, Inc.
- Product
- Water Pik Sinusense battery operated device including a blue semi-transparent water reservoir, a white hand held button initiated injector device, and a soft seal nozzle affixed to the white injector. The product is sold individually in white, blue red and orange cartons with a photo of the device as well as a stream of water flowing through it. The contents of the container include the device as well as a package insert with utilization instructions. Product Usage: Sinus washing device to keep sinus passages clear of congestion.
- Product code
- KMA Irrigator, Powered Nasal
- Reason for recall
- Customer complaints related to cross contamination of sinus rinse solution between the battery compartment and solution reservoir used for rinsing nasal passages.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Water Pik has generated a Press Release letter dated May 2012. The letter included safety information and instructions for the consumer to contact Waterpik for a replacement model if they experience water intrusion into the battery compartment of their device. For additional information go to www.waterpiksinusense.com/exchange.
- Quantity in commerce
- 105257 individual units
- Distribution
- US Nationwide Distribution.
- Product codes and lots
- All lots and codes manufactured between May 2010 and July 2011.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.