Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1673-2012

HPBLUE: HARMONIC Blue Hand Piece; HP054: HARMONIC Hand Piece; Non-Sterile products are individually packed in...

Ethicon Endo-Surgery Inc / initiated 2012-03-16 / Terminated / root cause: manufacturing process

Ethicon Endo-Surgery Inc began a recall of HPBLUE: HARMONIC Blue Hand Piece; HP054: HARMONIC Hand Piece; Non-Sterile products are individually packed in a poly bag and then placed into an individual... on . The reason FDA records is "Ethicon Endo-Surgery initiated a voluntary recall due to some lot numbers of the HARMONIC Hand Piece (HP054) and the HARMONIC Blue Hand Piece (HPBLUE), which impacts their compatibility with the HARMONIC Generator 300...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1673-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ethicon Endo-Surgery Inc
Product
HPBLUE: HARMONIC Blue Hand Piece; HP054: HARMONIC Hand Piece; Non-Sterile products are individually packed in a poly bag and then placed into an individual carton. Mfr. By: Ethicon Endo-Surgery, LLc, 475 Calle C, Guaynbo, PR 00969. Indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels.
Product code
LFL Instrument, Ultrasonic Surgical
Reason for recall
Ethicon Endo-Surgery initiated a voluntary recall due to some lot numbers of the HARMONIC Hand Piece (HP054) and the HARMONIC Blue Hand Piece (HPBLUE), which impacts their compatibility with the HARMONIC Generator 300 (GEN04). The problem occurs after successful set up and initial use of HARMONIC Hand Pieces (HP054 and HPBLUE) with the Generator 300 (GEN04). The procedure is interrupted when
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Ethicon Endo-Surgery Inc sent an URGENT: Product Correction Notification letter dated March 2012 to all affected customers. The letter identified the product, the problem, the action to be taken by the customer. Labels were enclosed that are used to identify the recalled product. Customers were instructed to fill out the Business Reply Form and fax it to Stericycle at 1-888-912-7091 within three business days for a credit. For questions regarding this recall call 513-337-3419.
Quantity in commerce
1896 instruments
Distribution
Worldwide Distribution-USA including: AK,AL,AR,AZ,CA,CO,CT,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NCND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI,WV, and WY. and the countries of: Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Ecuador, Egypt, Greece, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Korea (South),Kuwait, Lebanon, Malaysia, Mexico, Netherlands, Antilles, Pakistan, Panama, Puerto Rico, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Taiwan, Thailand, Turkey, United Arab Emirates Uruguay, Venezuela & Yemen.
Product codes and lots
HARMONIC¿ HP054 Hand Pieces Lot Numbers produced to-date: H92Y3M, J9039C, J90817, J90E7C, J9000C, J90432, J9084W, J90E7Z, J90029, J90470, J9088L, J90F3G,J9006C, J9054E, J9093L, J90F7Z, J9011D, J9056K, J9095C, J90G32, J90151, J9057V,J90974, J90G9Y,J9018H, J9058L, J90A45, J90H37,J9020F, J90624, J90A72, J90H4U, 9023D, J90681, J90C7X, J90H7D,J9027C, J9069F, J90C84, J90H9Y,J90314, J90726, J90D4N, J90J2N,J9034N, J90762, J90E0R, J90J8E,J9038W, J9076A, and J90E29.   HARMONIC¿ HPBLUE Hand Pieces Lot Numbers produced to-date: J9001P, J9053N, J9091G, J90D84,J9005J, J9057E, J90A2C, J90E55,J9010E, J9065U, J90A64, J90F76,J9042G, J90696, J90C36, J90G4C,J9050E, J9086W, J90C7U, and J90H46.