Advanced Neuromodulation Systems, Brio DBS System implantable neurostimulator, IPG or R-IPG, Model 6788...

Abbott Medical / initiated 2012-03-28 / Terminated / root cause: design

Advanced Neuromodulation Systems Inc. began a recall of Advanced Neuromodulation Systems, Brio DBS System implantable neurostimulator, IPG or R-IPG, Model 6788 Implantable Pulse Generator on . FDA classified it as Class II and gives the reason as "Unintended Fluid ingress into the device header has been shown". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0894-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Advanced Neuromodulation Systems, Brio DBS System implantable neurostimulator, IPG or R-IPG, Model 6788 Implantable Pulse Generator.
Product code
MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor
Reason for recall
Unintended Fluid ingress into the device header has been shown.
Root cause tag
design
FDA root cause
Device Design
Action
St. Jude Medical sent an Important Product Field Corrective Action Device Information letter dated March 28, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory for any unused product, isolate the affected product for return to St. Jude Medical. Their local representatiave would collect the unused inventory. Any unused product would be replaced with product once inventory becomes available. For patients with implanted product and no degradation of therapy, it is recommended dthat the devices not be unnecesssarily explanted. For patient with implanted product who experience diminished or loss of therapy customers were advised to follow the steps provided in the letter. Customers were asked to forward the notification to any institution in the event one or more patients or products could be potentially affected by this action. Customers wsere asked to maintain a record of thenotice along with the recommendations to ensure the effectiveness of the communication. For questions regarding this recall call 972-309-2154.
Quantity in commerce
388 units
Distribution
Worldwide Distribution - 53 consignees out of the US.
Product codes and lots
Model number 6788. All lots manufactured prior to 03/16/2012 are being recalled.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0894-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0894-2013Class IITerminatedVoluntary: Firm initiated