Z-0894-2013Class II
Advanced Neuromodulation Systems, Brio DBS System implantable neurostimulator, IPG or R-IPG, Model 6788...
Abbott Medical / initiated 2012-03-28 / Terminated / root cause: design
Advanced Neuromodulation Systems Inc. began a recall of Advanced Neuromodulation Systems, Brio DBS System implantable neurostimulator, IPG or R-IPG, Model 6788 Implantable Pulse Generator on . FDA classified it as Class II and gives the reason as "Unintended Fluid ingress into the device header has been shown". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0894-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Advanced Neuromodulation Systems, Brio DBS System implantable neurostimulator, IPG or R-IPG, Model 6788 Implantable Pulse Generator.
- Product code
- MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor
- Reason for recall
- Unintended Fluid ingress into the device header has been shown.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- St. Jude Medical sent an Important Product Field Corrective Action Device Information letter dated March 28, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory for any unused product, isolate the affected product for return to St. Jude Medical. Their local representatiave would collect the unused inventory. Any unused product would be replaced with product once inventory becomes available. For patients with implanted product and no degradation of therapy, it is recommended dthat the devices not be unnecesssarily explanted. For patient with implanted product who experience diminished or loss of therapy customers were advised to follow the steps provided in the letter. Customers were asked to forward the notification to any institution in the event one or more patients or products could be potentially affected by this action. Customers wsere asked to maintain a record of thenotice along with the recommendations to ensure the effectiveness of the communication. For questions regarding this recall call 972-309-2154.
- Quantity in commerce
- 388 units
- Distribution
- Worldwide Distribution - 53 consignees out of the US.
- Product codes and lots
- Model number 6788. All lots manufactured prior to 03/16/2012 are being recalled.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0894-2013 | Class II | Terminated | Voluntary: Firm initiated |