Z-1494-2012

Siemens-branded ONCOR" or PRIMUS" linear accelerator with an OPTIVUE" or MVision" imaging system; Model...

Siemens Medical Solutions USA, Inc. / initiated 2012-04-04 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Siemens-branded ONCOR" or PRIMUS" linear accelerator with an OPTIVUE" or MVision" imaging system; Model Numbers: ONCOR Expression 7360717, ONCOR Impression... on . The reason FDA records is "The firm became aware of a potential issue with Siemens-branded ONCOR or PRIMUS linear accelerator with an OPTIVUE or MVision imaging system that has the potential collision problem of the imaging panel if the panel...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1494-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens-branded ONCOR" or PRIMUS" linear accelerator with an OPTIVUE" or MVision" imaging system; Model Numbers: ONCOR Expression 7360717, ONCOR Impression Plus 5857912, ONCOR Avant Garde 5863472, PRIMUS HI 4504200; ARTISTE/ONCOR/PRIMUS systems are manufactured by: SIEMENS AG, MEDICAL SOLUTIONS, Roentgenstrasse 19-21, Kemnath, GERMANY 95478 Product Usage: Medical charged-particle radiation therapy system. The intended use of the SIEMENS branded ONCOR and PRIMUS family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
The firm became aware of a potential issue with Siemens-branded ONCOR or PRIMUS linear accelerator with an OPTIVUE or MVision imaging system that has the potential collision problem of the imaging panel if the panel does not retract before the gantry rotates; this may result in patient injury.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Siemens sent an Update Instruction containing the Customer Information letters beginning April 4, 2012 delivered by Siemens customer Service Group or by certified mail to all affected customers. The letter identified the affected product, what is the issue and what has Siemens done to address this issue. The letter instructs customer to include this Customer Information in their Digital Linear Accelerator System Owner Manual chapter Safety Advisory Letters where it should remain.
Quantity in commerce
73 units
Distribution
Worldwide Distribution - USA Nationwide
Product codes and lots
All units of these models.