Z-1493-2012

ARTISTE or ONCOR Linear Accelerators in combination with 160MLC (Multi Leaf Collimator). Linear Accelerators...

Siemens Medical Solutions USA, Inc. / initiated 2012-04-03 / Terminated / root cause: labeling and UDI

Siemens Medical Solutions USA, Inc began a recall of ARTISTE or ONCOR Linear Accelerators in combination with 160MLC (Multi Leaf Collimator). Linear Accelerators that may be affected are: ARTISTE MV System; ONCOR... on . The reason FDA records is "Siemens became aware of a potential safety risk related to the behavior of the 160 MLC when using the "Motion Stop" button followed by "Reset" several times during the treatment of one beam or one segment (issue 1)...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1493-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ARTISTE or ONCOR Linear Accelerators in combination with 160MLC (Multi Leaf Collimator). Linear Accelerators that may be affected are: ARTISTE MV System; ONCOR Avant Garde ONCOR Impression ONCOR Impression Plus ONCOR Expression. Manufactured by Siemens AG, Medical Solutions, Kemnath, Germany; Registered by Siemens Medical Solutions, Concord, CA. ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. Product Model with Part Number: ARTISTE MV System, 8139789; ONCOR Avant Garde, 5863472; ONCOR Impression, 5857920; ONCOR Impression, Plus 5857912; ONCOR Expression, 7360204; ONCOR Expression, 7360717. ARTISTE and ONCOR family of Linear Accelerators are used to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Siemens became aware of a potential safety risk related to the behavior of the 160 MLC when using the "Motion Stop" button followed by "Reset" several times during the treatment of one beam or one segment (issue 1). Additionally, there may be a potential for collision if the operator does not ensure clearance of the extended flat panel positioner for all gantry rotation angles (issue 2).
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Siemens sent a Safety Advisory Notice with Update to all affected customers. The notice informed the customers of the risk of pressing the Motion Stop button repeatedly during treatment. Work Steps were provided for customers to follow.
Distribution
Worldwide Distribution - USA including the states of CA, DE, FL, IL, KY, LA, MA, MI, NE, NJ, NY, OH, PA, SD, TN, UT, WI, and WV
Product codes and lots
Product Model with Part Number: ARTISTE MV System, 8139789; ONCOR Avant Garde, 5863472; ONCOR Impression, 5857920; ONCOR Impression, Plus 5857912; ONCOR Expression, 7360204; ONCOR Expression, 7360717