Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1497-2012

MicroScan Dried Overnight Gram Negative Panels; Panel type: Neg Urine Combo 55; Catalog Number B1017-409...

Siemens Healthcare Diagnostics Inc. / initiated 2011-06-16 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of MicroScan Dried Overnight Gram Negative Panels; Panel type: Neg Urine Combo 55; Catalog Number B1017-409. Siemens Healthcare Diagnostics, West Sacramento, CA on . The reason FDA records is "Three lots were confirmed to not meet performance claims due to random non-biological contamination which is only apparent (precipated) after inoculation of the panels". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1497-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
MicroScan Dried Overnight Gram Negative Panels; Panel type: Neg Urine Combo 55; Catalog Number B1017-409. Siemens Healthcare Diagnostics, West Sacramento, CA.
Product code
JWY Manual Antimicrobial Susceptibility Test Systems
Reason for recall
Three lots were confirmed to not meet performance claims due to random non-biological contamination which is only apparent (precipated) after inoculation of the panels.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Customer Notification letters were sent to all customers of the three MicroScan panels by traceable mail. The letter identified the affected product and asked customers to discard any remaining inventory in addition to contacting customer service for replacement product. The letter also recommended that customers consider the need to review previous test results, conduct patient follow-up, and or repeat tests as needed. Customers were to complete and return the customer reply card to confirm that they received and understood the recall communication.
Quantity in commerce
687 boxes (20 panels per box).
Distribution
Worldwide Distribution -- US, Turkey, Poland, and Great Britain.
Product codes and lots
lot code 2011-06-09.