MicroScan Dried Overnight Gram Negative Panels; Panel type: Neg Breakpoint Combo Type 47, catalog number...
Siemens Healthcare Diagnostics, Inc. began a recall of MicroScan Dried Overnight Gram Negative Panels; Panel type: Neg Breakpoint Combo Type 47, catalog number: B1017-417. Siemens Healthcare Diagnostics, West... on . The reason FDA records is "Three lots were confirmed to not meet performance claims due to random non-biological contamination which is only apparent (precipated) after inoculation of the panels". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1499-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- MicroScan Dried Overnight Gram Negative Panels; Panel type: Neg Breakpoint Combo Type 47, catalog number: B1017-417. Siemens Healthcare Diagnostics, West Sacramento, CA. For use in determining quantitative and / or qualitative antimicrobial agent susceptibility and / or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-negative bacteria.
- Product code
- JWY Manual Antimicrobial Susceptibility Test Systems
- Reason for recall
- Three lots were confirmed to not meet performance claims due to random non-biological contamination which is only apparent (precipated) after inoculation of the panels
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Customer Notification letters were sent to all customers of the three MicroScan panels by traceable mail. The letter identified the affected product and asked customers to discard any remaining inventory in addition to contacting customer service for replacement product. The letter also recommended that customers consider the need to review previous test results, conduct patient follow-up, and or repeat tests as needed. Customers were to complete and return the customer reply card to confirm that they received and understood the recall communication.
- Quantity in commerce
- 683 boxes (20 panels per box)
- Distribution
- Worldwide Distribution -- US, Turkey, Poland, and Great Britain.
- Product codes and lots
- lot code 2011-12-15.