Dimension Vista(R) Total Prostate Specific Antigen Flex(R) reagent cartridge
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista(R) Total Prostate Specific Antigen Flex(R) reagent cartridge Product Usage: The TPSA method is an in vitro diagnostic test for the quantitative... on . The reason FDA records is "Siemens has confirmed a report of leaking Flex(R) reagent cartridge lot 11300BB from Dimension Vista(R) TPSA; when a leaking Flex(R) is used, patient test results may be falsely depressed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1540-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista(R) Total Prostate Specific Antigen Flex(R) reagent cartridge Product Usage: The TPSA method is an in vitro diagnostic test for the quantitative measurement of total prostate specific antigen (PSA) in human serum and plasma on the Dimension Vista(R) System
- Product code
- MTF Total, Prostate Specific Antigen (Noncomplexed & Complexed) For Detection Of Prostate Cancer
- Reason for recall
- Siemens has confirmed a report of leaking Flex(R) reagent cartridge lot 11300BB from Dimension Vista(R) TPSA; when a leaking Flex(R) is used, patient test results may be falsely depressed.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens sent an Urgent Field Safety Notification letter dated April 14, 2012 to all Dimension Vista(R) System customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately discontinue use and discard any remaining inventory of the affected product. Customers were asked to complete the attached form and fax it to 302-631-8467 to indicate that you have received the notification. For technical questions or concerns, contact the Siemens Technical Solutions Center at 800-441-9250.
- Quantity in commerce
- 1,103 cartons
- Distribution
- USA Nationwide Distribution including the states of: AL, AZ, CA, CO, CT, FL IL, LA, MD, MI, MN, MS, MT, NC, NH, NJ, NV, NY, OH, PA, SD, TN, TX, UT, WI ,WV including Puerto Rico.
- Product codes and lots
- Catalog number: K6451; Mnemonic: TPSA; lot number 11300BB, exp 10/26/2012