Terumo Custom Cardiovascular Procedure Kit (manufactured to the design and specifications of the...
Terumo Cardiovascular Systems Corp began a recall of Terumo Custom Cardiovascular Procedure Kit (manufactured to the design and specifications of the purchaser/user) for the application it is intended by the... on . The reason FDA records is "Loose connections (i.e., connections between PVC circuit tubing and Oxygenator ports) may result in leaks and patient blood loss". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1523-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Custom Cardiovascular Procedure Kit (manufactured to the design and specifications of the purchaser/user) for the application it is intended by the user. X-COATED CHANGE OUT PACK WITH FX15E Product Code: 74257
- Product code
- DTZ Oxygenator, Cardiopulmonary Bypass
- Reason for recall
- Loose connections (i.e., connections between PVC circuit tubing and Oxygenator ports) may result in leaks and patient blood loss
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Terumo Cardiovascular Systems issued an Urgent Medical Device Recall notification on 4/12/12 to all user facilities/consignees who have received affected devices. The letter identified the affected product, stated the reason for the recall, the potential hazard involved, and the correction. Users are to review the letter, ensure that all users are made aware of the issue, discontinue use of the affected product, and return it to Terumo CVS. In addition, users are to return the response form to the manufacturer. Affected units will be replaced with new units. Contact Terumo CVS Customer Service @ 1-800-521-2818 or [email protected]
- Quantity in commerce
- 3 units
- Distribution
- Worldwide Distribution -- US, including the states of AL, AZ, CA, MO, and NY and the country of Canada.
- Product codes and lots
- Lot No.:PA30