Z-1500-2012

AtheNA Multi-Lyte MMV Test System- A93111G A microparticle-based immunoassay intended for the qualitative...

Zeus Scientific LLC / initiated 2012-03-08 / Terminated / root cause: manufacturing process

Zeus Scientific, Inc. began a recall of AtheNA Multi-Lyte MMV Test System- A93111G A microparticle-based immunoassay intended for the qualitative presumptive detection of IgG class antibody to the... on . The reason FDA records is "Select lots of AtheNA Multi-Lyte MMV Test System have lost reactivity and may produce false positive results in seronegative patients who have not been previously affected with or vaccinated against Mumps". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1500-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zeus Scientific LLC
Product
AtheNA Multi-Lyte MMV Test System- A93111G A microparticle-based immunoassay intended for the qualitative presumptive detection of IgG class antibody to the Mumps virus in human serum using the AtheNA Multi-Lyte Test System. The test system is intended to be used for determination of a previous infection with the Mumps virus.
Product code
LJY Enzyme Linked Immunoabsorbent Assay, Mumps Virus
Reason for recall
Select lots of AtheNA Multi-Lyte MMV Test System have lost reactivity and may produce false positive results in seronegative patients who have not been previously affected with or vaccinated against Mumps.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Zeus Scientific, sent an "URGENT NOTICE; Voluntary Product Recall" letter dated March 8, 2012 their customers via Fed Ex. The letter described the product, problem and actions to be taken. The customers were instructed to remove and discard any remaining inventory of the listed product; if you are not the end-user and/or you have distributed these products to customers, notify them accordingly; and complete, sign and return the verification notice via fax to: 908-526-2058 or email a PDF to [email protected]. If you have any questions, call 908-526-3744 or 800-286-2111.
Quantity in commerce
180 kits
Distribution
Nationwide distribution.
Product codes and lots
11120087 11120165  11110125 : (This is an amendment to add the additional lot 11110125 which was left off.. This amendment was made on 6/6/2012.)