Z-1482-2012

Cardiac Resynchronization Therapy (CRT) Leads. Trade Name: QuickSite, QuickSite XL Model Numbers: QuickSite...

Abbott Medical / initiated 2012-04-03 / Terminated / root cause: design

St Jude Medical CRMD began a recall of Cardiac Resynchronization Therapy (CRT) Leads. Trade Name: QuickSite, QuickSite XL Model Numbers: QuickSite and QuickSite XL: Models 1056T, model 1058T... on . The reason FDA records is "Visual observations of externalized conductors due to abrasion of the silicone insulation in the distal portion of these QuickSite and QuickFlex leads". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1482-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Cardiac Resynchronization Therapy (CRT) Leads. Trade Name: QuickSite, QuickSite XL Model Numbers: QuickSite and QuickSite XL: Models 1056T, model 1058T Intended use: Left Ventricular Cardiac Resynchronization Therapy Lead
Product code
NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Reason for recall
Visual observations of externalized conductors due to abrasion of the silicone insulation in the distal portion of these QuickSite and QuickFlex leads.
Root cause tag
design
FDA root cause
Device Design
Action
A recall communication was initiated on 4/3/12 with St. Jude Medical forwarding a Medical Device Advisory to all doctors who have received the St. Jude QuickSite and QuickFlex Left Ventricular CRT leads models 1056T, 1058T, 1156T, and 1158T. The letter provides the doctors with the problems identified and actions to be taken. Doctors with any questions or concerns are instructed to contact their local St. Jude Medical representative or their Technical Services Department at (800) 722-3774. A Physician Device Advisory Notice for the QuickSite and QuickFlex Left Ventricular CRT Leads was issued on 4/3/12 to physicians who have received the recalled products. The notice provides the physicians with the problem identified and actions to be taken. Physicians are instructed to return any packaged product in hospital inventory to SJM Sales Representatives. For implanted product, due to the nature of the issue, externalized conductors may be observed with no electrical or performance anomalies. Therefore, prophylactic explant is not recommended. It is recommended that physicians continue to monitor their patients' implanted system at regularly scheduled intervals. A Medical Advisory and Inventory Management Information letter was sent on 4/3/12 to hospital administrators to notify them of the problems identified and the actions to be taken.
Quantity in commerce
101,000 units (59,000 units in US)
Distribution
Worldwide Distribution -- Algeria, Argentina, Australia, Austria, Bahrain, Balkan, Baltic, Belarus, Belgium, Bolivia, Bosnia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Mauritius, Montenegro, Morocco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Paraguay, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Senegal, Serbia, Serbia Montenegro, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, US, Venezuela, and Yemen.
Product codes and lots
All lots