Z-1658-2012

Becton Dickinson Phoenix Update Disks (PUD), catalog # 441107/ 448047, component of BD automated microbiology...

Becton Dickinson & Co / initiated 2012-04-10 / Terminated / root cause: manufacturing process

Becton Dickinson & Co. began a recall of Becton Dickinson Phoenix Update Disks (PUD), catalog # 441107/ 448047, component of BD automated microbiology system, and used with BD Gram Positive MIC or... on . The reason FDA records is "BD identified an increase in vancomycin-resistant strains that may exhibit erroneous results on all PMIC or PMIC/ID panels and could lead to ineffective treatment regimen in affected patients". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1658-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
Becton Dickinson Phoenix Update Disks (PUD), catalog # 441107/ 448047, component of BD automated microbiology system, and used with BD Gram Positive MIC or MIC/ID panels. Product Usage: The Phoenix Update Disk (PUD) delivers substrate and drug specific information to be used by the BD Phoenix instrument software in determining organism identification and susceptibility results. Information contained in the PUD does not change instrument functionality or features.
Product code
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Reason for recall
BD identified an increase in vancomycin-resistant strains that may exhibit erroneous results on all PMIC or PMIC/ID panels and could lead to ineffective treatment regimen in affected patients.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Becton Dickinson (BD) sent an Urgent Field Correction Action letter dated April 2012 and an Update letter dated April 18, 2012 to all affected customers via e-mail and UPS overnight delivery. The letter identified the affected products, problem and actions to be taken. The letter provided specific workaround instructions for device users pending software upgrades. A BD Service Engineer will contact affected customers to arrange for the installation of the updated software when it becomes available. The letter included an Acknowledgement Form to be completed. For questions contact Technical Services at 800-638-8663.
Quantity in commerce
3,454,993 total products (U.S.)
Distribution
Worldwide Distribution - USA (nationwide) including the states of: AL, AR, AZ, CA, CO, CT ,DC ,FL , GA, IA, IL, IN, KY, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, AND WY. and the countries of Canada, Australia, Philippines, Thailand, Malaysia, India, New Zealand, Belgium, Singapore, Japan, Taiwan, China, Chile, Colombia, Korea, Brazil and Venezuela.
Product codes and lots
BD Phoenix Update Disks (PUD) versions: 4.51 R, 4.56G, 4.61A, 4.65A, 4.71A, 4.75A, 4.81A, 4.85A, 4.91A, 5.11A, 5.15A, 5.21A