Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1508-2012

Product is labeled in part: Inner Label "***SMART Control Transhepatic Biliary Stent***REF Cat. No.***Lot...

Cordis US Corp. / initiated 2012-04-09 / Terminated / root cause: packaging

Cordis Corporation began a recall of Product is labeled in part: Inner Label "***SMART Control Transhepatic Biliary Stent***REF Cat. No.***Lot No.***Use By***Cordis S.M.A.R.T. CONTROL Nitinol... on . The reason FDA records is "Cordis Corporation is recalling 32 lots of Cordis S.M.A.R.T. Nitinol Stent Transhepatic Biliary System and S.M.A.R.T. Control Nitinol Stent Transhepatic Biliary System. There is a potential for a sterility breach due to...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1508-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cordis US Corp.
Product
Product is labeled in part: Inner Label "***SMART Control Transhepatic Biliary Stent***REF Cat. No.***Lot No.***Use By***Cordis S.M.A.R.T. CONTROL Nitinol Stent Transhepatic Biliary System***Assembled in Mexico***Cordis Corporation 14201 North West 60th Avenue Miami Lakes, Florida 33014, USA***Cordis S.M.A.R.T. CONTROL BILIARY***Cordis a Johnson & Johnson company***Endovascular***" Catalog C10080MB Lot No. 15535360. Catalog C10080SB Lot No. 15530747. Catalog C12030SB Lot No. 15530756. Catalog C12040SB Lot No. 15530755. Catalog C12060SB Lot No's. 15535363, 15569685. Catalog C12080MB Lot No. 15530757 Catalog C14040SB Lot No's. 15530754, 15557313. Catalog C14060SB Lot No's. 15530753, 15557311. Catalog C14080SB Lot No. 15535361. Indicated for palliation of malignant neoplasms in the biliary tree.
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
Cordis Corporation is recalling 32 lots of Cordis S.M.A.R.T. Nitinol Stent Transhepatic Biliary System and S.M.A.R.T. Control Nitinol Stent Transhepatic Biliary System. There is a potential for a sterility breach due to holes or tears in the Tyvek packaging.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Cordis Corp sent a " Urgent Medical Device Recall (Removal) letter dated April 9, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. A Response Form was provided with each notification. Customers were asked to complete and return Response Form. The customers were instructed to : " Immediately identify and set aside all product listed above in a manner that ensures the affected product will not be used. " Review, complete, sign and return the enclosed Acknowledgement Form directly to Cordis at the fax number on the form. " Either return any affected product with reference to the RA# on the form, or contact your local sales representative, or contact Customer Service (see Available Assistance section) to facilitate return of the affected product. Replacement product will be provided in a short time frame. Credit will be provided for customer-owned inventory upon request. " Pass on this notice to anyone in your facility that needs to be informed. " If any product listed below has been forwarded to another facility, contact that facility to arrange return. " Maintain awareness of this notice until all product listed above has been returned to Cordis. For questions related to the issue or product replacement, please contact your local sales representative, or contact the special Customer Service number (800) 551-7683, Monday through Friday from 8:30 AM to 8:00 PM EDT.
Quantity in commerce
278 units
Distribution
Worldwide Distribution-- USA (nationwide) and the country Canada.
Product codes and lots
Lot No's. 15535360, 15530747, 15530756, 15530755, 15535363, 15569685, 15530757, 15530754, 15557313, 15530753, 15557311, 15535361.