Integra Mozaik Strip 10 cc
Integra LifeSciences Corp. began a recall of Integra Mozaik Strip 10 cc Product Usage: The product is a bone graft substitute composed of collagen (Type 1) and ceramic (I3-Tricalcium phosphate) matrix in... on . FDA classified it as Class II and gives the reason as "The Instructions for Use {IFU} {Putty Part Number 7700713001 and Strip Part Number 770151300l for CE marked Mozaik Strip and Putty products contains an incorrect translation of the term "bone marrow aspirate". The term...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1795-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Integra LifeSciences Corporation
- Product
- Integra Mozaik Strip 10 cc Product Usage: The product is a bone graft substitute composed of collagen (Type 1) and ceramic (I3-Tricalcium phosphate) matrix in putty and strip forms. The product is intended for use as a bone void filler to fill voids or gaps of the skeletal system in the extremities, spine. and pelvis not intrinsic to the stability of the bony structure. Integra Mozaik is also indicated for use in the treatment of surgically treated osseous defects or osseous defects created from traumatic injury to the bone. Following placement in the bone void or a defect Integra Mozaik is reabsorbed and replaced with bone during the healing process.
- Product code
- MQV Filler, Bone Void, Calcium Compound
- Reason for recall
- The Instructions for Use {IFU} {Putty Part Number 7700713001 and Strip Part Number 770151300l for CE marked Mozaik Strip and Putty products contains an incorrect translation of the term "bone marrow aspirate". The term is incorrectly translated in Spanish and French,
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Integra LifeSciences sent a Field Safety Notice letter dated October 11, 2011 to their sole foreign consignee. The letter identified the affected products, problem and actions to be taken. The consignee was asked to remove the existing Instructions for Use (IFU) and replace it with the corrected IFU. The consignee was instructed to complete and return the attached Distributor Rework and Acknowledgment Forms provided with the customer notification letter as documentation that instructions were followed and all required actions were completed. For any questions, call: +33 (0)4 37 47 5900 , fax: +33 (0)4 37 47 5999; or email: [email protected].
- Quantity in commerce
- 22 units
- Distribution
- Worldwide Distribution
- Product codes and lots
- Strip Part Number 7701513001 510 K- K063124 Device Listing Number - D001098 Integra MOZAIKTM Strip IOcc Reference number - CCM I021OITL Batch Number - 105BR0215955 Integra MOZAIKTM Strip 10cc Reference Number - CCM I021OITL Batch Number - 105BR0216149
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1795-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1796-2012 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.