QuietCare, Model Numbers QC200403, v.02; QC200403, v.03; QC200403, v.04; QC200405 (USA) and QCE200403, v.02...
Intel-GE Care Innovations LLC began a recall of QuietCare, Model Numbers QC200403, v.02; QC200403, v.03; QC200403, v.04; QC200405 (USA) and QCE200403, v.02; QCE200403, v.03 (UK). Manufacturer Living... on . The reason FDA records is "The QuietCare production system experienced a partial outage on 7/29/11. As a result of the partial outage, some monitoring data about residents' movements was not processed in the production system, and alerts were...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1615-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intel-Ge Care Innovations, LLC
- Product
- QuietCare, Model Numbers QC200403, v.02; QC200403, v.03; QC200403, v.04; QC200405 (USA) and QCE200403, v.02; QCE200403, v.03 (UK). Manufacturer Living Independently Group, Inc., New York, NY. QuietCare is intended for use in monitoring the environmental conditions and activity (motion) of an individual living in a senior housing community.
- Product code
- IQA System, Environmental Control, Powered
- Reason for recall
- The QuietCare production system experienced a partial outage on 7/29/11. As a result of the partial outage, some monitoring data about residents' movements was not processed in the production system, and alerts were therefore not sent to caregivers. The problem was identified and corrected and the unprocessed data was recovered and processed. However, because of the time lag in processing, a fe
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Initial notification to the affected facilities was done by telephone or email on 7/30/11. This was followed up with Urgent Field Notification letters to the customers on 8/12/11 via e-mail. For questions, customers were instructed to contact Customer Support at 1 (866) 772-8243.
- Quantity in commerce
- 82 units affected by the outage: Domestic - 80 units, Foreign - 2 units
- Distribution
- Worldwide distribution - USA and the United Kingdom.
- Product codes and lots
- Affected Serial Numbers: liqc.018778, liqc.020447, liqc.020428, liqc.024472, liqc.027039, liqc.024195, liqc.026718, liqc.023495, liqc.027065, liqc.024252, liqc.018722, liqc.024377, liqc.019808, liqc.017029, liqc.017310, liqc.026701, liqc.024190, liqc.019075, liqc.027351, liqc.023253, liqc.019497, liqc.019215, liqc.019217, liqc.017965, liqc.020530, liqc.024293, liqc.021530, liqc.026353, liqc.017900, liqc.019915, liqc.018447, liqc.021086, liqc.027072, liqc.023241, liqc.018169, liqc.017876, liqc.022049, liqc.026409, liqc.018126, liqc.019519, liqc.017579, liqc.018271, liqc.017531, liqc.017532, liqc.018255, liqc.023888, liqc.023340, liqc.023343, liqc.023342, liqc.018534, liqc.018536, liqc.018256, liqc.017532, liqc.019582, liqc.021064, liqc.020981, liqc.021944, liqc.021764, liqc.021284, liqc.017741, liqc.017738, liqc.024298, liqc.026760, liqc.027063, liqc.022321, liqc.022320, liqc.020843, liqc.022009, liqc.026620, liqc.026643, liqc.026456, liqc.021428, liqc.021402, liqc.022054, liqc.019577, liqc.023782, liqc.023710, liqc.023708, liqc.019360, liqc.017682, liqc.021670, liqc.022983 (UK), liqc.022846 (UK).