Z-1535-2012

Terumo Custom Cardiovascular Procedure Kit- X-COATED PERFUSION PK FX15RE40

Terumo Cardiovascular Systems Corporation / initiated 2012-04-12 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corp began a recall of Terumo Custom Cardiovascular Procedure Kit- X-COATED PERFUSION PK FX15RE40 Product Code:73673 Manufactured to the design and specifications of the... on . The reason FDA records is "Loose connections (i.e., connections between PVC circuit tubing and Oxygenator ports) may result in leaks and patient blood loss". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1535-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Custom Cardiovascular Procedure Kit- X-COATED PERFUSION PK FX15RE40 Product Code:73673 Manufactured to the design and specifications of the purchaser/user) for the application it is intended.
Product code
DTZ Oxygenator, Cardiopulmonary Bypass
Reason for recall
Loose connections (i.e., connections between PVC circuit tubing and Oxygenator ports) may result in leaks and patient blood loss
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Terumo Cardiovascular Systems issued an Urgent Medical Device Recall notification on 4/12/12 to all user facilities/consignees who have received affected devices. The letter identified the affected product, stated the reason for the recall, the potential hazard involved, and the correction. Users are to review the letter, ensure that all users are made aware of the issue, discontinue use of the affected product, and return it to Terumo CVS. In addition, users are to return the response form to the manufacturer. Affected units will be replaced with new units. Contact Terumo CVS Customer Service @ 1-800-521-2818 or [email protected]
Quantity in commerce
7 units
Distribution
Worldwide Distribution -- US, including the states of AL, AZ, CA, MO, and NY and the country of Canada.
Product codes and lots
Lot No: NA17