Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1886-2012Class II

X-Coated Fx25rw H/l Pack

Terumo Cardiovascular Systems Corporation / initiated 2012-05-09 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of X-COATED FX25RW H/L PACK Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to... on . FDA classified it as Class II and gives the reason as "During production of the Sarns Antegrade Cardioplegia Cannula, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of a foreign substance on the inner surface of some cannulae tips. Terumo CVS'...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1886-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
X-COATED FX25RW H/L PACK Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during cardiopulmonary bypass procedures.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
During production of the Sarns Antegrade Cardioplegia Cannula, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of a foreign substance on the inner surface of some cannulae tips. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface. The substance was likely deposited during the molding process, but the exact composition of the
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Terumo sent an Urgent Medical Device Recall letter dated May 9, 2012 to All US first consignees will receive a removal notification letter via Federal Express. The letter identified the affected products, reason for correction and actions to be taken. Customers were instructed to discontinue use of affected products and return for replacement. The letter advised customers to assure that all users are aware of this notice and to fill out and return the attached Response Form. For questions call Terumo CVS Customer Service at 1-800-521-2818 or by fax at 1-800-292-6551.
Quantity in commerce
53,139 (each) in total
Distribution
Worldwide Distribution - US (nationwide) including the states of: AL, AR , AZ , CA, CT , FL, IA , IL, IN , KS , LA , MI , NM, MO , NC, NE , NJ , NV, NY , OH, OK , PA, RI , SC , TN, TX, WA , WI, WV and the countries of: BELGIUM, BRAZIL, Byelorussian SSR, CAMBODIA , Dominican Republic, Honduras, Japan, Malaysia, Mexico, Nicaragua, Saudi Arabia, Singapore, South Africa, Thailand, Trinidad and Tobago, UNITED ARAB EMIRATES (UAE), and Vietnam.
Product codes and lots
Catalog number: 74301 and lot number: PD19

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1886-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1875-2012Class IITerminatedVoluntary: Firm initiated
Z-1876-2012Class IITerminatedVoluntary: Firm initiated
Z-1877-2012Class IITerminatedVoluntary: Firm initiated
Z-1878-2012Class IITerminatedVoluntary: Firm initiated
Z-1879-2012Class IITerminatedVoluntary: Firm initiated
Z-1880-2012Class IITerminatedVoluntary: Firm initiated
Z-1881-2012Class IITerminatedVoluntary: Firm initiated
Z-1882-2012Class IITerminatedVoluntary: Firm initiated
Z-1883-2012Class IITerminatedVoluntary: Firm initiated
Z-1884-2012Class IITerminatedVoluntary: Firm initiated
Z-1885-2012Class IITerminatedVoluntary: Firm initiated
Z-1886-2012Class IITerminatedVoluntary: Firm initiated
Z-1887-2012Class IITerminatedVoluntary: Firm initiated
Z-1888-2012Class IITerminatedVoluntary: Firm initiated
Z-1889-2012Class IITerminatedVoluntary: Firm initiated
Z-1890-2012Class IITerminatedVoluntary: Firm initiated
Z-1891-2012Class IITerminatedVoluntary: Firm initiated
Z-1892-2012Class IITerminatedVoluntary: Firm initiated
Z-1893-2012Class IITerminatedVoluntary: Firm initiated
Z-1894-2012Class IITerminatedVoluntary: Firm initiated