Z-1813-2012Class III
PrepStain Installation Kit (PreCoat), Catalog/Model # 490528
Becton Dickinson & Co / initiated 2012-03-22 / Terminated / root cause: design
Tripath Imaging, Inc. began a recall of PrepStain Installation Kit (PreCoat), Catalog/Model # 490528 on . FDA classified it as Class III and gives the reason as "The PrepStain Syringing Pipettes may not dispense the minimum volume of 7.5 mL of the sample due to leaking. This event resulted in an increase in complaints of leaking syringing pipettes". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1813-2012
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- PrepStain Installation Kit (PreCoat), Catalog/Model # 490528
- Product code
- MKQ Processor, Cervical Cytology Slide, Automated
- Reason for recall
- The PrepStain Syringing Pipettes may not dispense the minimum volume of 7.5 mL of the sample due to leaking. This event resulted in an increase in complaints of leaking syringing pipettes
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- BD Diagnostics sent a "BD PrepMate Syringing Pipette" letter dated March 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The firm encouraged customers to refer to the Operator's Manual and inspect the centrifuge tubes to verify that the correct amount of sample has been transferred. Customers experiencing this issue were instructed to discard the pipettes and contact BD Diagnostic Technical Support for a replacement. Contact the firm at 1-877-822-7771 for questions regarding this notice.
- Quantity in commerce
- 46,112,784 for all products in Recall Event
- Distribution
- Worldwide Distribution.
- Product codes and lots
- 100917ry, 101105sh, 110107rg
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1811-2012 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-1812-2012 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-1813-2012 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-1814-2012 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-1815-2012 | Class III | Terminated | Voluntary: Firm initiated |