Z-1813-2012Class III

PrepStain Installation Kit (PreCoat), Catalog/Model # 490528

Becton Dickinson & Co / initiated 2012-03-22 / Terminated / root cause: design

Tripath Imaging, Inc. began a recall of PrepStain Installation Kit (PreCoat), Catalog/Model # 490528 on . FDA classified it as Class III and gives the reason as "The PrepStain Syringing Pipettes may not dispense the minimum volume of 7.5 mL of the sample due to leaking. This event resulted in an increase in complaints of leaking syringing pipettes". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1813-2012
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
PrepStain Installation Kit (PreCoat), Catalog/Model # 490528
Product code
MKQ Processor, Cervical Cytology Slide, Automated
Reason for recall
The PrepStain Syringing Pipettes may not dispense the minimum volume of 7.5 mL of the sample due to leaking. This event resulted in an increase in complaints of leaking syringing pipettes
Root cause tag
design
FDA root cause
Component design/selection
Action
BD Diagnostics sent a "BD PrepMate Syringing Pipette" letter dated March 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The firm encouraged customers to refer to the Operator's Manual and inspect the centrifuge tubes to verify that the correct amount of sample has been transferred. Customers experiencing this issue were instructed to discard the pipettes and contact BD Diagnostic Technical Support for a replacement. Contact the firm at 1-877-822-7771 for questions regarding this notice.
Quantity in commerce
46,112,784 for all products in Recall Event
Distribution
Worldwide Distribution.
Product codes and lots
100917ry, 101105sh, 110107rg

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1813-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1811-2012Class IIITerminatedVoluntary: Firm initiated
Z-1812-2012Class IIITerminatedVoluntary: Firm initiated
Z-1813-2012Class IIITerminatedVoluntary: Firm initiated
Z-1814-2012Class IIITerminatedVoluntary: Firm initiated
Z-1815-2012Class IIITerminatedVoluntary: Firm initiated