Z-1849-2012Class II

Medtronic 0-arm Imaging System with Software Version 3.1.1, Catalog Numbers: BI-700-00027-120R and...

Medtronic Navigation, Inc / initiated 2010-10-12 / Terminated / root cause: software

Medtronic Navigation, Inc. began a recall of Medtronic 0-arm Imaging System with Software Version 3.1.1, Catalog Numbers: BI-700-00027-120R and BI-700-00027-120. The 0-arm Imaging System is a mobile x-ray... on . FDA classified it as Class II and gives the reason as "Software Version 3.1.1 does not meet navigational accuracy". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1849-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic 0-arm Imaging System with Software Version 3.1.1, Catalog Numbers: BI-700-00027-120R and BI-700-00027-120. The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging.
Product code
IZL System, X-Ray, Mobile
Reason for recall
Software Version 3.1.1 does not meet navigational accuracy.
Root cause tag
software
FDA root cause
Software design
Action
Medtronic Navigation contacted customers by telephone on 10/12/2010 and with "Urgent Field Safety Notice" follow up letters dated 10/14 or 10/15/2010. Sites were visited by Medtronic Service Representative to complete navigational accuracy testing and either replace or correct the unit.
Quantity in commerce
3 units
Distribution
Nationwide Distribution including the states of CO, IN, and MN.
Product codes and lots
Serial Numbers: 125, 156, 234R.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1849-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1849-2012Class IITerminatedVoluntary: Firm initiated