Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1836-2012Class II

SARNS FLEXIBLE ARTERIAL CANNULA: 7.3 MM (22 FR) 00 WITH 3/8" CONNECTOR,SUTURE RING, 9.5" (24 CM) LONG. The...

Terumo Cardiovascular Systems Corporation / initiated 2012-05-09 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of SARNS FLEXIBLE ARTERIAL CANNULA: 7.3 MM (22 FR) 00 WITH 3/8" CONNECTOR,SUTURE RING, 9.5" (24 CM) LONG. The Sarns High Flow Aortic Arch Cannula is indicated for... on . FDA classified it as Class II and gives the reason as "During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1836-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
SARNS FLEXIBLE ARTERIAL CANNULA: 7.3 MM (22 FR) 00 WITH 3/8" CONNECTOR,SUTURE RING, 9.5" (24 CM) LONG. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the parti
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Terumo sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 9, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter advises customers to discontinue use of the affected product and to return the product to the firm. A Response Form was attached for customers to complete and return. Contact Terumo CVS Customer Service at 1-800-521-2818 for questions regarding this recall.
Quantity in commerce
133,793 (each)
Distribution
Worldwide Distribution-USA (nationwide) including the states of AL , AR , AZ, CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , MD , ME , MI , MI , MN , MO , MT , NC , ND , NE , NH , NJ , NM , NV , NY , OH , OK , OR , PA , RI , SC , SD , TN , TX , UT , VA , WA , WI, and WY and the countries of Australia, Belgium, Brazil, Canada, Chile, Germany, Honduras, Hong Kong, India, Japan, Mauritius, Nicaragua, Philippines, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates (UAE), and Vietnam.
Product codes and lots
Catalog number: 13020 and Lot codes: 0550552, 0555854, 0557327, 0559103, 0562503, 0566861, 0570455, 0574279, 0577885, 0581606, 0585963, 0590856, 0593457, 0597308, 0605179, 0609199, 0614248, 0618969, 0626379, 0628950, 0634171, 0636332, 0640518, 0646396, 0653231, and 0666719.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1836-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1832-2012Class IITerminatedVoluntary: Firm initiated
Z-1833-2012Class IITerminatedVoluntary: Firm initiated
Z-1834-2012Class IITerminatedVoluntary: Firm initiated
Z-1835-2012Class IITerminatedVoluntary: Firm initiated
Z-1836-2012Class IITerminatedVoluntary: Firm initiated
Z-1837-2012Class IITerminatedVoluntary: Firm initiated
Z-1838-2012Class IITerminatedVoluntary: Firm initiated
Z-1839-2012Class IITerminatedVoluntary: Firm initiated
Z-1840-2012Class IITerminatedVoluntary: Firm initiated
Z-1841-2012Class IITerminatedVoluntary: Firm initiated
Z-1842-2012Class IITerminatedVoluntary: Firm initiated