SPECIALTYCARE - HUDGENS. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the...
Terumo Cardiovascular Systems Corporation began a recall of SPECIALTYCARE - HUDGENS. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery on . FDA classified it as Class II and gives the reason as "During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1840-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- SPECIALTYCARE - HUDGENS. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.
- Product code
- DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
- Reason for recall
- During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the parti
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Terumo sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 9, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter advises customers to discontinue use of the affected product and to return the product to the firm. A Response Form was attached for customers to complete and return. Contact Terumo CVS Customer Service at 1-800-521-2818 for questions regarding this recall.
- Quantity in commerce
- 133,793 (each)
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AL , AR , AZ, CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , MD , ME , MI , MI , MN , MO , MT , NC , ND , NE , NH , NJ , NM , NV , NY , OH , OK , OR , PA , RI , SC , SD , TN , TX , UT , VA , WA , WI, and WY and the countries of Australia, Belgium, Brazil, Canada, Chile, Germany, Honduras, Hong Kong, India, Japan, Mauritius, Nicaragua, Philippines, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates (UAE), and Vietnam.
- Product codes and lots
- Catalog number: 64625 and Lot code: NA17, NC07A, NG06, NG06A, and NL19.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1832-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1833-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1834-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1835-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1836-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1837-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1838-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1839-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1840-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1841-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1842-2012 | Class II | Terminated | Voluntary: Firm initiated |