Z-1700-2012
IMMULITE --- Insulin --- Catalog Number: LKIN1 (100 test), LKIN5 (500 tests) --- Test Code: INS, Color: Orange
Siemens Healthcare Diagnostics Inc. / initiated 2012-04-30 / Terminated
Siemens Healthcare Diagnostics began a recall of IMMULITE --- Insulin --- Catalog Number: LKIN1 (100 test), LKIN5 (500 tests) --- Test Code: INS, Color: Orange Product Usage: For in vitro diagnostic use with... on . The reason FDA records is "Siemens Healthcare Diagnostics has observed a negative bias in the Immulite Insulin assay". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1700-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- IMMULITE --- Insulin --- Catalog Number: LKIN1 (100 test), LKIN5 (500 tests) --- Test Code: INS, Color: Orange Product Usage: For in vitro diagnostic use with IMMULITE and IMMULITE 1000 Systems Analyzers --- for the quantitative measurement of insulin in serum or heparinized plasma, for the management of diabetes.
- Product code
- CFP Radioimmunoassay, Immunoreactive Insulin
- Reason for recall
- Siemens Healthcare Diagnostics has observed a negative bias in the Immulite Insulin assay.
- FDA root cause
- Component change control
- Action
- Siemens sent an Urgent Device Recall letter dated April 2012 via fax to affected U.S. customers and an Urgent Field Safety Notice to customers outside the U.S. The letter identified the affected kit lots, reason for voluntary recall, risk to health and actions to be taken by the customer/user. Customers were instructed to discard any inventory of the affected kit lots and complete and return the attached Field Correction Effectiveness Check questionnaire within thirty (30) days to request a no-charge replacement kits. Customers were requested to post this field correction by each IMMULITE system to alert users to the situation. For questions contact your Siemens Technical Solutions Center.
- Distribution
- Worldwide Distribution - US (nationwide)
- Product codes and lots
- kit lot 0335