Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1701-2012

IMMULITE 2000 --- Insulin --- Catalog Number: L2KIN2 (200 test), L2KIN6 (600 tests) --- Test Code: INS...

Siemens Healthcare Diagnostics Inc. / initiated 2012-04-30 / Terminated

Siemens Healthcare Diagnostics began a recall of IMMULITE 2000 --- Insulin --- Catalog Number: L2KIN2 (200 test), L2KIN6 (600 tests) --- Test Code: INS, Color: Orange --- Product Usage: For in vitro... on . The reason FDA records is "Siemens Healthcare Diagnostics has observed a negative bias in the Immulite Insulin assay". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1701-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
IMMULITE 2000 --- Insulin --- Catalog Number: L2KIN2 (200 test), L2KIN6 (600 tests) --- Test Code: INS, Color: Orange --- Product Usage: For in vitro diagnostic use with IMMULITE and IMMULITE 1000 Systems Analyzers --- for the quantitative measurement of insulin in serum or heparinized plasma, for the management of diabetes.
Product code
CFP Radioimmunoassay, Immunoreactive Insulin
Reason for recall
Siemens Healthcare Diagnostics has observed a negative bias in the Immulite Insulin assay.
FDA root cause
Component change control
Action
Siemens sent an Urgent Device Recall letter dated April 2012 via fax to affected U.S. customers and an Urgent Field Safety Notice to customers outside the U.S. The letter identified the affected kit lots, reason for voluntary recall, risk to health and actions to be taken by the customer/user. Customers were instructed to discard any inventory of the affected kit lots and complete and return the attached Field Correction Effectiveness Check questionnaire within thirty (30) days to request a no-charge replacement kits. Customers were requested to post this field correction by each IMMULITE system to alert users to the situation. For questions contact your Siemens Technical Solutions Center.
Distribution
Worldwide Distribution - US (nationwide)
Product codes and lots
kit lots 312, 313, 314, 315, 316