Z-1504-2012
Caridian BCT COBE Spectra System Blood Warmer, Catalog number 950000000
Terumo BCT, Ltd / initiated 2011-10-26 / Terminated / root cause: design
CaridianBCT, Inc. began a recall of Caridian BCT COBE Spectra System Blood Warmer, Catalog number 950000000 on . The reason FDA records is "Potential for air entry into the fluid pathway of the COBE Spectra due to a leak at the return luer connection to an elevated bloodwarmer when directions for use are not followed". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1504-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo BCT, Ltd
- Product
- Caridian BCT COBE Spectra System Blood Warmer, Catalog number 950000000.
- Product code
- LKN Separator, Automated, Blood Cell And Plasma, Therapeutic
- Reason for recall
- Potential for air entry into the fluid pathway of the COBE Spectra due to a leak at the return luer connection to an elevated bloodwarmer when directions for use are not followed.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Caridian BCT sent a letter to customers dated June 2011. It was a Field labeling addendum to the COBE Spectra Apheresis System Operator's Manual stating: "When connecting a blood warmer tubing set to the return line, ensure that the tubing connection is tight. Put the luer connection no higher than 20 in (50 cm) above the return access to prevent the possibility of air entering the tubing." This correction was distributed to all consignees and is included with the current version of the owner's manual.
- Quantity in commerce
- 2459 machines
- Distribution
- Worldwide distribution including USA, Canada, Algeria, Argentina, Australia, Austria, Bahamas, Belgium, Brazil, Chile, China, Colombia, Czech Republic, Denmark, Netherlands, UK, Hong Kong, Hungary, India, Italy, Iran, Ireland, Israel, Jordan, Korea, Kuwait, Lebanon, Libya, Luxembourg, Egypt, Finland, France, Germany, Japan, Malaysia, Mexico, Norway, Oman, Pakistan, Philippines, Poland, Portugal, Romania, Russia, Spain, Saudi Arabia, South Africa, Sweden, Switzerland, Taiwan, Thailand, Ukraine, Uruguay, Vatican, Vietnam, and Greece.
- Product codes and lots
- All serial numbers.