Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1706-2012

Stryker Xia 3 Anti-Torque Key Manual Surgical Instrument; Catalog Number 48237026

Howmedica Osteonics Corp / initiated 2012-01-12 / Terminated / root cause: manufacturing process

Stryker Spine began a recall of Stryker Xia 3 Anti-Torque Key Manual Surgical Instrument; Catalog Number 48237026. Product Usage: : Reusable instrument used to aid in the final tightening of... on . The reason FDA records is "Stryker received four reports regarding a handle which separates at the point where it is welded to the shaft on the Xia 3 Anti-Torque Key during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1706-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Xia 3 Anti-Torque Key Manual Surgical Instrument; Catalog Number 48237026. Product Usage: : Reusable instrument used to aid in the final tightening of the spinal construct. Intended Use: The Anti-Torque Key allows the Torque Wrench to align with the tightening axis and maximizing the torque needed to lock the implant assembly during final tightening of the spinal construct.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
Stryker received four reports regarding a handle which separates at the point where it is welded to the shaft on the Xia 3 Anti-Torque Key during surgery.
Root cause tag
manufacturing process
FDA root cause
Mixed-up of materials/components
Action
Stryker Spine sent an Urgent Product Recall letter dated January 12, 2012 with an attached Customer Response forms via FEDEX to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine their inventory to identify the affected product and return using the prepaid mailing label provided. The letter asked customers to complete the Customer Response Form and fax to 201.760.8370. For questions call 201.760.8298.
Quantity in commerce
95 units
Distribution
US Nationwide Distribution
Product codes and lots
Lot Codes (510k exempt): 07F116, 07F117, 085525, 086577, 086578, 087232, 087233, 088604.