Stryker Xia 3 Anti-Torque Key Manual Surgical Instrument; Catalog Number 48237026
Stryker Spine began a recall of Stryker Xia 3 Anti-Torque Key Manual Surgical Instrument; Catalog Number 48237026. Product Usage: : Reusable instrument used to aid in the final tightening of... on . The reason FDA records is "Stryker received four reports regarding a handle which separates at the point where it is welded to the shaft on the Xia 3 Anti-Torque Key during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1706-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Xia 3 Anti-Torque Key Manual Surgical Instrument; Catalog Number 48237026. Product Usage: : Reusable instrument used to aid in the final tightening of the spinal construct. Intended Use: The Anti-Torque Key allows the Torque Wrench to align with the tightening axis and maximizing the torque needed to lock the implant assembly during final tightening of the spinal construct.
- Product code
- NKB Thoracolumbosacral Pedicle Screw System
- Reason for recall
- Stryker received four reports regarding a handle which separates at the point where it is welded to the shaft on the Xia 3 Anti-Torque Key during surgery.
- Root cause tag
- manufacturing process
- FDA root cause
- Mixed-up of materials/components
- Action
- Stryker Spine sent an Urgent Product Recall letter dated January 12, 2012 with an attached Customer Response forms via FEDEX to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine their inventory to identify the affected product and return using the prepaid mailing label provided. The letter asked customers to complete the Customer Response Form and fax to 201.760.8370. For questions call 201.760.8298.
- Quantity in commerce
- 95 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Lot Codes (510k exempt): 07F116, 07F117, 085525, 086577, 086578, 087232, 087233, 088604.