Z-1956-2012Class II

CIVCO VirtuTRAX Instrument Navigator, Reusable non-sterile general purpose driveBAY, trakSTAR...

CIVCO Medical Instruments Co., Inc. / initiated 2012-05-07 / Terminated / root cause: design

Civco Medical Instruments Inc began a recall of CIVCO VirtuTRAX Instrument Navigator, Reusable non-sterile general purpose driveBAY, trakSTAR, electromagnetic sensor (8mm) with VirtuTRAX Instrument navigator... on . FDA classified it as Class II and gives the reason as "The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1956-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CIVCO Medical Instruments Co., Inc.
Product
CIVCO VirtuTRAX Instrument Navigator, Reusable non-sterile general purpose driveBAY, trakSTAR, electromagnetic sensor (8mm) with VirtuTRAX Instrument navigator and (8.9 tapered to 1.3 x 147cm) telescopically-folded CIV-Flex sensor covers (5), Ref. 610-1080. The firm name on the label is CIVCO, Kalona, IA. The kit contains sensor component part number 410-240. The general purpose sensor within the kit is intended to provide physicians with tools for electromagnetic tracking of instruments with respect to image data. The sensors are designed to attach to custom reusable tracking bracket for leading transducers and other medical devices
Product code
IYO System, Imaging, Pulsed Echo, Ultrasonic
Reason for recall
The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.
Root cause tag
design
FDA root cause
Component design/selection
Action
CIVCO sent an Urgent Medical Device Recall letter dated May 3, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to return all general purpose sensors as the product may not function as intended. To return the affected product customers should contact the OEM Sales Coordinator at 319-248-6521 to schedule the return of their device. Upon receipt of the snesor, CIVCO will arrange for recalibration. Once recalibration is completed, the sensor will be returned to the customer at no charge. Customers were asked to forward this communication to all their customers who purchased the affected product. Once CIVCO has received the customer list, they will contact the end-user to request they return their general purpose sensor directly to CIVCO for recalibration. For questions customers should call 800-441-6741. For questions regarding this recall call 319-656-4447.
Quantity in commerce
24 kits
Distribution
Worldwide Distribution - USA including WI, CA, MA, IN, WA, and NJ. Foreign distribution was made to Germany, Canada, Denmark, Hong Kong, Ukraine, Japan, Italy, Korea, and France.
Product codes and lots
Lot numbers M493010, M493220, M708150, M710060, M783360, M803040, M803520, M808870, M823690, M832260, M840620, M848450, M855680, and M871810.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1956-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1952-2012Class IITerminatedVoluntary: Firm initiated
Z-1953-2012Class IITerminatedVoluntary: Firm initiated
Z-1954-2012Class IITerminatedVoluntary: Firm initiated
Z-1955-2012Class IITerminatedVoluntary: Firm initiated
Z-1956-2012Class IITerminatedVoluntary: Firm initiated