Accessories for the Captiva Spine SmartLOX Cervical Plate System. Products are labeled in part: "***Part...
Captiva Spine, Inc began a recall of Accessories for the Captiva Spine SmartLOX Cervical Plate System. Products are labeled in part: "***Part Number CP40012***12mm Drill Bit***Lot Number 89427-2... on . The reason FDA records is "Captiva Spine is recalling the SmartLOX Cervical Plate components 12 & 14 mm Drill Bits, Single Barrel Variable Drill Guides, and Double Barrel Fixed Drill Guides. The components are part of the instrumentation kit used...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1648-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CAPTIVA SPINE, Inc.
- Product
- Accessories for the Captiva Spine SmartLOX Cervical Plate System. Products are labeled in part: "***Part Number CP40012***12mm Drill Bit***Lot Number 89427-2 R***" "***Part Number CP40014***14mm Drill Bit***Lot Number 89427-3-R***" The Captiva Spine SmartLOX Cervical Plate System is intended for the anterior screw fixation of the cervical spine. These implants have been designed to provide stabilization as an adjunct to cervical fusion.
- Product code
- KWQ Appliance, Fixation, Spinal Intervertebral Body
- Reason for recall
- Captiva Spine is recalling the SmartLOX Cervical Plate components 12 & 14 mm Drill Bits, Single Barrel Variable Drill Guides, and Double Barrel Fixed Drill Guides. The components are part of the instrumentation kit used to complete an anterior screw fixation of the cervical spine. There is a potential risk for the drills to not properly fit through the drill guides due to an overlap in the dime
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Captiva Spine notified customers of the recall by letter on November 23, 2011, and advised that the affected products would be replaced. Customers were asked to return the drill guides and unused drills with the provided return labels. For questions customers should call 877-772-5571. For questions regarding this recall call 561-277-9480.
- Quantity in commerce
- 212 units
- Distribution
- Nationwide Distribution including TX, IN, UT, AZ, GA, CA, AL, MI, and MN.
- Product codes and lots
- Lot Numbers: 89427-2 R & 89427-3-R