Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1655-2012

Vascular Solutions, GuideLiner V2 Catheter, Rx only, Sterile EO, Catalog Number 5571, 6F, Vascular Solutions...

Vascular Solutions LLC / initiated 2012-01-17 / Terminated / root cause: manufacturing process

Vascular Solutions, Inc. began a recall of Vascular Solutions, GuideLiner V2 Catheter, Rx only, Sterile EO, Catalog Number 5571, 6F, Vascular Solutions, Inc., 6464 Sycamore Court, Minneapolis, MN 55369... on . The reason FDA records is "Vascular Solutions has became aware of a potential defect with their 6F GuideLiner V2 catheters (Model 5571) with lot number 555038. Use of this product could result in the catheter shaft separating from the pushwire on...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1655-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Vascular Solutions LLC
Product
Vascular Solutions, GuideLiner V2 Catheter, Rx only, Sterile EO, Catalog Number 5571, 6F, Vascular Solutions, Inc., 6464 Sycamore Court, Minneapolis, MN 55369. Intended to be used in conjunction with guide catheters to access discrete regions of the coronary and or peripheral vasculature and too facilitate placement and exchange of guidewired and other interventional devices.
Product code
DQY Catheter, Percutaneous
Reason for recall
Vascular Solutions has became aware of a potential defect with their 6F GuideLiner V2 catheters (Model 5571) with lot number 555038. Use of this product could result in the catheter shaft separating from the pushwire on the device. This may cause a procedural delay and/or injury possibly requiring medical intervention.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Vascular Solutions sent an "URGENT: MEDICAL DEVICE RECALL" letter dated January 17, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to check their inventory for the affected product , immediately remove the product and place the product in a secure area. Additionally, customers were asked to complete and return a Field Action Customer Inventory Form to the Customer Service Department via fax at 763-656-4251 or e-mail at [email protected]. Customers may contact their local sales representative for questions concerning this recall.
Quantity in commerce
330
Distribution
Nationwide Distribution including the states of AL, AZ, AR, CO, CT, DC, FL, GA, HI, IL, IN, IA, KS, KY, LAM ME, MD, MA, MI, MN, MS, MO, MT, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, WA, WV, and WI.
Product codes and lots
555038