Z-1538-2012Class II

Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide

Philips North America LLC / initiated 2011-10-19 / Terminated

Philips Healthcare Inc. began a recall of Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide on . FDA classified it as Class II and gives the reason as "When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1538-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide
Product code
IZL System, X-Ray, Mobile
Reason for recall
When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when component supplie
FDA root cause
Radiation Control for Health and Safety Act
Action
Philips Healthcare sent a proprosed corrective action letter dated May 3, 2012, to FDA. The proposed corrective action plan is to bring the product into compliance with regulatory requirements by issuing a field corrective action which should be completed by Philips by August 1, 2012. For questions regarding this recall call 978-687-1501.
Quantity in commerce
11 - US
Distribution
Nationwide Distribution
Product codes and lots
n/a

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1538-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1538-2012Class IITerminatedVoluntary: Firm initiated