Z-1703-2012

Brand Name: QuietCare, Model Numbers QC200403, v.02; QC200403, v.03; QC200403, v.04 -- Classification Name...

Intel-Ge Care Innovations, LLC / initiated 2011-08-01 / Terminated / root cause: manufacturing process

Intel-GE Care Innovations LLC began a recall of Brand Name: QuietCare, Model Numbers QC200403, v.02; QC200403, v.03; QC200403, v.04 -- Classification Name: System, Environmental Control, Powered. -- Firm on... on . The reason FDA records is "On 7/29/11, the QuietCare production system experienced a partial outage from about 15:00 hours until 22:30. During the engineering investigation into the cause of the service disruption, it was discovered that some...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1703-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intel-Ge Care Innovations, LLC
Product
Brand Name: QuietCare, Model Numbers QC200403, v.02; QC200403, v.03; QC200403, v.04 -- Classification Name: System, Environmental Control, Powered. -- Firm on label: Manufacturer Living Independently Group, Inc., A wholly owned subsidiary of General Electric Company, PO Box 1466, New York, NY 10116 USA, 1-866-772-8243 --- The system is a passive behavior-monitoring device. QuietCare uses strategically placed, unobtrusive, passive wireless infrared pyroelectric sensors in the senior's residence to detect their movement. Data from these sensors are sent to a base station that periodically transmits the information to a centralized secure file server through standard phone lines over a toll-free number. Sophisticated algorithms analyze the data, learning each individual's normal patterns of behavior. When the system detects significant deviations from these baselines it automatically alerts designated caregivers. QuietCare also generates alerts of late wake-ups, potential bathroom falls and if household temperatures become dangerously low or high. The behavioral changes that the system tracks can often serve as subtle indicators of the potential emerging health problems. These changes include: nighttime bathroom usage, entering or exiting the bedroom at a different time than usual, change in the number of times the refrigerator door is opened/closed change in the number of times the medicine cabinet door is opened/closed. The intended users of the QuietCare device are the caregivers at the Assisted Living Facilities and Independent Living Facilities. The monitored individuals (seniors) do not actively operate the device and the device function is not dependent upon the senior resident's ability to use the device. --- Class 2 device, 510(k) exempt, Device Listing Number D117156.
Product code
IQA System, Environmental Control, Powered
Reason for recall
On 7/29/11, the QuietCare production system experienced a partial outage from about 15:00 hours until 22:30. During the engineering investigation into the cause of the service disruption, it was discovered that some residents did not have the wander and night motion features properly activated, and alerts were not being transmitted. Failure to generate alerts in a timely manner could result in a
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Intel-GE Care Innovations sent an "URGENT FIELD NOTIFICATION" e-mail dated August 12, 2011 to all affected customers. The notification identifies the product and problem. No action is required by the customers. Contact Customer Service at (866) 772-8243 for questions regarding this notice.
Quantity in commerce
61 units were identified as affected
Distribution
Nationwide Distribution-USA (nationwide) including the states of Arizona, California, Connecticut, Idaho, Illinois, Minnesota, New Jersey, New York, Ohio, Pennsylvania, Texas, and Utah.
Product codes and lots
Affected Hardware Serial Numbers: 020685, 020267, 021542, 018880, 021585, 024179, 022003, 020232, 020272, 024148, 024252, 017044, 017648, 023385, 018651, 024376, 023988, 018567, 017151, 018460, 026701, 021130, 021366, 017735, 019229, 026371, 026317, 018010, 026334, 017471, 019952, 026293, 019476, 026292, 017089, 019953, 026335, 018630, 026439, 026378, 026260, 026239, 022050, 019477, 018379, 024346, 018224, 023936, 020421, 022129, 022216, 018543, 026760, 024270, 021649, 021918, 023465, 023531, 020716, 018962,  020718.