Philips Easy Diagnost
Philips Healthcare Inc. began a recall of Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system on . FDA classified it as Class II and gives the reason as "If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1821-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system
- Product code
- IXR Table, Radiographic, Tilting
- Reason for recall
- If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Philips sent an Urgent - Field Safety Notice letter dated May 11, 2012 to their affected customers. The letter identified the affected product, what the issues are, under what circumstances they may occur and the actions to be taken to avoid or minimize the occurrence of the issue Philips plans to supply an Addendum to the Instruction for Use. A Philips Service Engineer will contact the affected customers as soon as the Field Action Kit is ready to be implemented. For questions contact your local Philips representative: Technical Support Line 1-800-722-9377.
- Quantity in commerce
- 6
- Distribution
- 6 US customers
- Product codes and lots
- All systems with Eleva software version 4.x and stitching option.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1821-2012 | Class II | Terminated | Voluntary: Firm initiated |