Z-1821-2012Class II

Philips Easy Diagnost

Philips North America LLC / initiated 2012-05-11 / Terminated / root cause: design

Philips Healthcare Inc. began a recall of Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system on . FDA classified it as Class II and gives the reason as "If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1821-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system
Product code
IXR Table, Radiographic, Tilting
Reason for recall
If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut
Root cause tag
design
FDA root cause
Device Design
Action
Philips sent an Urgent - Field Safety Notice letter dated May 11, 2012 to their affected customers. The letter identified the affected product, what the issues are, under what circumstances they may occur and the actions to be taken to avoid or minimize the occurrence of the issue Philips plans to supply an Addendum to the Instruction for Use. A Philips Service Engineer will contact the affected customers as soon as the Field Action Kit is ready to be implemented. For questions contact your local Philips representative: Technical Support Line 1-800-722-9377.
Quantity in commerce
6
Distribution
6 US customers
Product codes and lots
All systems with Eleva software version 4.x and stitching option.