Rectal Retractors, stainless steel, item # AL07289000, a component of Varian Medical Ring and Tandem...
Varian Medical Systems, Inc. began a recall of Rectal Retractors, stainless steel, item # AL07289000, a component of Varian Medical Ring and Tandem Applicator Sets; Intended for use in Brachytherapy on . FDA classified it as Class II and gives the reason as "A medical device used in brachytherapy may become contaminated and can cause infections, toxicity or allergic reactions". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0759-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- Rectal Retractors, stainless steel, item # AL07289000, a component of Varian Medical Ring and Tandem Applicator Sets; Intended for use in Brachytherapy.
- Product code
- JAQ System, Applicator, Radionuclide, Remote-Controlled
- Reason for recall
- A medical device used in brachytherapy may become contaminated and can cause infections, toxicity or allergic reactions.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Varian Medical Systems an an Urgent Medical Device Correction letters dated April 11, 2012 and May 8, 2012 to all affected customers. The April notification advised of an anomaly with titanium rectal retractors, and the correction was expanded in May to include stainless steel retractors. In both notifications, users were instructed to discontinue use and quarantine their cleaned and sterilized devices for collection by Varian field service engineers. Users were requested to return an enclosed reply form to respond with number of units in inventory.
- Quantity in commerce
- 344 pieces
- Distribution
- US Nationwide Distribution
- Product codes and lots
- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0759-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0760-2013 | Class II | Terminated | Voluntary: Firm initiated |