Z-1869-2012Class II

Siemens Dimension Vista(R) System B12 Flex(R) Reagent Cartridge (K6442) B12 Flex Reagent Cartridge used with...

Siemens Healthcare Diagnostics Inc. / initiated 2012-05-08 / Terminated / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Dimension Vista(R) System B12 Flex(R) Reagent Cartridge (K6442) B12 Flex Reagent Cartridge used with the Dimension Vista System for the in vitro... on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics has confirmed that the presence of Intrinsic Factor Blocking Antibody in some patients can result in a false and significant elevation in the reported concentration of Vitamin B12 on the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1869-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens Dimension Vista(R) System B12 Flex(R) Reagent Cartridge (K6442) B12 Flex Reagent Cartridge used with the Dimension Vista System for the in vitro diagnostic test for the quantitative measurement of Vitamin B12 in human serum and plasma.
Product code
CDD Radioassay, Vitamin B12
Reason for recall
Siemens Healthcare Diagnostics has confirmed that the presence of Intrinsic Factor Blocking Antibody in some patients can result in a false and significant elevation in the reported concentration of Vitamin B12 on the Dimension Vista System. This issue affects all lots of B12 Flex reagent cartridges on the Dimension Vista System.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens Healthcare Diagnostics Inc. sent a Urgent Device Recall letter to all affected customers dated May 2012. The letter identified the product the problem and the action needed to be taken by the customer. The letter instructed customers to immediately discontinue use of the device, quarantine and dispose of all inventory, ensure that your Medical Director reviews this letter as well as the attached physician letter. Please complete the attached Field Correction form and fax to 302-631-8467. Further questions please call 302-631-6311.
Quantity in commerce
59,735
Distribution
Worldwide Distribution -- USA (nationwide) and the countries of Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Great Britian, Italy, Netherlands, Norway, Portugal, Slovakia, South Africa, Spain, Switzerland, United Arab Emirates, Canada, New Zealand, and South Korea. *** Center Recommended Depth is Retail***
Product codes and lots
B12 Flex(R) reagent cartridge (K6442) - IFBA Interference All lots -- 07304AA, Exp 5/28/08; 08021AB, Exp 1/20/09; 08042AC, Exp 2/10/09; 08071AA, Exp 3/11/09; 08114AC, Exp 4/23/09; 09019AC, Exp 1/19/10; 09019AB, Exp 1/19/10; 09040AA, Exp 2/10/10; 09068AA, Exp 3/9/10; 09085AB, Exp 3/26/10; 09110AE, Exp 4/20/10; 09140AD, Exp 5/19/10; 09180AC, Exp 6/29/10; 09215AA, Exp 8/3/10; 09236AC, Exp 8/24/10; 09279AC, Exp 10/6/10; 09307AA, Exp 11/3/10; 10012AA, Exp 1/13/11; 10039AA, Exp 2/5/11; 10060AB, Exp 3/1/11; 10074AB, Exp 3/16/11; 10097AB, Exp 4/7/11; 10109AD, Exp 4/19/11; 10123AA, Exp 5/3/11; 10138AB, Exp 5/18/11; 10155AC, Exp 6/3/11; 10181AA, Exp 7/1/11;  10208AE, Exp 7/27/11; 10208AD, Exp 7/27/11; 10221AC, Exp 8/9/11; 10237AD, Exp 8/25/11; 10243AE, Exp 8/31/11; 10265AB, Exp 9/22/11; 10281AB, Exp 10/8/11; 10301AC, Exp 10/28/11; 10322AB, Exp 11/18/11; 10341AC, Exp 12/6/11; 10341AD, Exp 12/7/11; 11012AB, Exp 1/13/12; 11038AF, Exp 2/9/12;  11038AG, Exp 2/8/12; 11066AE, Exp 3/7/12; 11087AD, Exp 3/27/12; 11118AB, Exp 4/28/12; 11145AA, Exp 5/24/12; 11171AA, Exp 6/21/12; 11234AA, Exp 8/21/12; 11234AB, Exp 8/21/12; 11250AA, Exp 9/6/12; 11263AA, Exp 9/19/12; 11283AA, Exp 10/9/12; 11297AB, Exp 10/23/12; 11312AA, Exp 11/6/12; 11332AA, Exp 11/27/12; 11346AA, Exp 12/11/12; 11361AB, Exp 12/27/12; 12011AA, Exp 1/15/13; 12030AB, Exp 1/30/13; 12044AB, Exp 2/13/12; 12059AB, Exp 3/27/13; 12079AB, Exp 3/18/13; and 12093AB, Exp 4/1/13.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1869-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1869-2012Class IITerminatedVoluntary: Firm initiated