Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1786-2012

Accessories for the Captiva Spine FuseLOX Lumbar Cage. The FuseLOX Lumbar Inserter product is a reuseable...

CAPTIVA SPINE, Inc. / initiated 2012-04-27 / Terminated / root cause: design

Captiva Spine, Inc began a recall of Accessories for the Captiva Spine FuseLOX Lumbar Cage. The FuseLOX Lumbar Inserter product is a reuseable component of the Captiva Spine FuseLOX Lumbar... on . The reason FDA records is "Captiva Spine, Inc. initiated a recall of FuseLOX Lumbar Implants, Trials, and Inserters after FDA advised the firm that there is inadequate support for implemented design changes to the cleared device Pivotec Lumbar...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1786-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CAPTIVA SPINE, Inc.
Product
Accessories for the Captiva Spine FuseLOX Lumbar Cage. The FuseLOX Lumbar Inserter product is a reuseable component of the Captiva Spine FuseLOX Lumbar product. It is not generally sold as an individually packaged product. Inserter products are shipped as a component of the FuseLOX Lumbar Caddy which is placed in the FuseLox Lumbar Sterilization tray. The sterilization tray is distributed on consignment. FPT - F is FuseLOX, P is Plif, and I is Inserter. Products subject to recall: FuseLOX Lumbar Inserter Inserter - Part Number: FPT0200 Lot Number 6110100. Quantity distributed (8) is found in table on page 5 of 6. The FuseLOX Lumbar Cage implant is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The FuseLOX Lumbar Cage Implant is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine.
Product code
MAX Intervertebral Fusion Device With Bone Graft, Lumbar
Reason for recall
Captiva Spine, Inc. initiated a recall of FuseLOX Lumbar Implants, Trials, and Inserters after FDA advised the firm that there is inadequate support for implemented design changes to the cleared device Pivotec Lumbar Interbody Fusion Device.
Root cause tag
design
FDA root cause
Finished device change control
Action
Captiva Spine sent a Voluntary Correction Notification letter dated May 4, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to locate and return affected devices. Customers were also instructed to complete and return a Tracking/Verification Form. For questions customers were instructed to call 877-772-5571. For questions regarding this recall call 561-277-9480.
Quantity in commerce
8 units
Distribution
USA (nationwide) including FL,TX, UT, and NJ.
Product codes and lots
Lot Number: 6110100.