Z-1936-2012Class II

Microbiologics 217 Osseo Avenue North, St. Cloud, MN 56303, Gram Stain Control Slides, SLO3-10, ATCC Licensed...

Microbiologics, Inc. / initiated 2012-05-29 / Terminated / root cause: manufacturing process

Microbiologics Inc began a recall of Microbiologics 217 Osseo Avenue North, St. Cloud, MN 56303, Gram Stain Control Slides, SLO3-10, ATCC Licensed Derivative, MediMart Europe, 11 rue Emile, Zola... on . FDA classified it as Class II and gives the reason as "Microbiologics is recalling a number of lots of KWIK-QC Gram Stain Slides. This product gram negative control organism (E.coli) is not properly fixated on the slide, which may cause the user to unintentionally...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1936-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microbiologics, Inc.
Product
Microbiologics 217 Osseo Avenue North, St. Cloud, MN 56303, Gram Stain Control Slides, SLO3-10, ATCC Licensed Derivative, MediMart Europe, 11 rue Emile, Zola B.P. 2332, 38933 Grenoble Cedex 2 France. They are microscope slide preparations contains specific organisms populations of known and predictable characteristics. These slide support formal quality assurance programs by serving as quality control challenges to demonstrate and document satisfactory performance of staining regents and methods used in staining procedures and to document personnel's ability to perform the procedure and to properly interpret the microscopic test results.
Product code
LJG Quality Control Slides
Reason for recall
Microbiologics is recalling a number of lots of KWIK-QC Gram Stain Slides. This product gram negative control organism (E.coli) is not properly fixated on the slide, which may cause the user to unintentionally contaminate the patient area of the slide by flooding the control well with stain. This contamination could potentially lead to misdiagnosis and improper treatment of a patient.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Microbiologics sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 30, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers not to use or distribute the affected product and to quarantine and return to the firm any remaining product. A Customer Response Form was attached for customers to complete and return to the firm. Contact Customer Service at 320-229-7057 for questions regarding this recall.
Quantity in commerce
13,720 slides (10 slides per box)
Distribution
Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, GA, IL, IA, KY, ME, MA, MI, MN, MS, NJ, NM, NY, PA, PR, SC, TX, VT, and VA. and the countries of SWITZERLAND, ITALY, ISRAEL, IRELAND, COLOMBIA, SWEDEN, INDIA, SINGAPORE, SERBIA, SAUDI ARABIA, CANADA, TAIWAN, SOUTH AFRICA, FRANCE, BRAZIL, MALTA, MACEDONIA, MALAYSIA, FINLAND, ECUADOR, VENEZUELA, REPUBLIC OF KOREA, NETHERLANDS, PERU, UNITED ARAB EMERITES, TURKEY, POLAND, and LEBANON.
Product codes and lots
0353 - 0355, 0357 - 0363

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1936-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1936-2012Class IITerminatedVoluntary: Firm initiated