Alaris PC unit model 8015
CareFusion 303, Inc. began a recall of Alaris PC unit model 8015 Product Usage: The device is labeled for prescription use only on . FDA classified it as Class I and gives the reason as "The recall was initiated because the Alaris PC unit model 8015 has a component on the PC unit power supply board that is causing an error code(120.4630) "System Error" or "Missing Battery Error" at start up. The error...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2076-2012
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- Alaris PC unit model 8015 Product Usage: The device is labeled for prescription use only.
- Product code
- FRN Pump, Infusion
- Reason for recall
- The recall was initiated because the Alaris PC unit model 8015 has a component on the PC unit power supply board that is causing an error code(120.4630) "System Error" or "Missing Battery Error" at start up. The error code is accompanied by both an audible alarm and visual error messages on the PC unit screen.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Carefusion sent an Urgent: Medical Device Recall Notification letter dated June 27, 2012 to all their customers who purchased the Alaris PC unit (model 8015). The letter identified the affected units, explanation of the problem, potential risk and actions to be taken. Customers were instructed to not return their devices. If customers observe either of the error messages identified in the letter, then they were instructed to remove the PC unit from use and contact the Carefusion Support Center. Carefusion will contact their facility by phone within 60 days of receiving this notification to schedule a visit to replace their affected power supply board on their PC unit. Customers were instructed to promptly complete and return the enclosed customer response card to expedite the corrective action process. For questions and support contact: CareFusion Support Center at 888-562-6018 for (Recall Related Questions), Customer Advocacy at 800-854-7128, Option 1, Option 1, Option 3 or email [email protected] for (Adverse Event Reports), Technical Support at 888-812-3229 for (Technical Questions Regarding the Alaris System).
- Quantity in commerce
- 112,912 units total (106,888 units in US)
- Distribution
- US Nationwide Distribution
- Product codes and lots
- not available
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2076-2012 | Class I | Terminated | Voluntary: Firm initiated |